Operator’s Manual Addendum<br />
Alarm setup<br />
Touch the ALARM SETUP button to display the current alarm settings and change<br />
the alarm settings as needed. A low circuit pressure (3P PEAK) alarm is now available<br />
during NIV to detect potential circuit disconnects or large system leaks based upon<br />
pressure measurements in the patient circuit. Refer to Table 4 and Table 6 <strong>for</strong> more<br />
in<strong>for</strong>mation regarding this alarm. The 3P PEAK alarm may be turned OFF, if desired.<br />
Figure 9 shows the NIV alarm screen with new patient default settings.<br />
NIV<br />
indicator<br />
4P PEAK<br />
alarm limit<br />
Figure 9. New patient alarm settings<br />
34 10070808 Rev. C Software Enhancements