Operator’s Manual Addendum<br />

Update to graphics<br />

This section updates Operator’s Manual section 6 to reflect these enhancements to<br />

graphics displays:<br />

• The PLOT 1 and PLOT 2 buttons now display drop-down menus of available<br />

selections with the current selection highlighted.<br />

• If a pressure-time wave<strong>for</strong>m is selected and the spontaneous type is TC or PA, a<br />

new SHADOW TRACE button is displayed. The SHADOW TRACE button<br />

allows you to enable or disable the graphic display of carinal pressure or lung<br />

pressure when TC or PA is active.<br />

• The pressure-time curve now shows an estimate of carinal pressure (PCARI) or<br />

lung pressure (PLUNG) as a shaded area within the wave<strong>for</strong>m when the<br />

spontaneous breath type is TC or PA, respectively.<br />

NOTE:<br />

The graphic display of carinal or lung pressure is an estimate, not an actual<br />

measurement.<br />

Printing graphics<br />

Frozen graphics can now be printed. When graphics are frozen on the screen and<br />

the selected device <strong>for</strong> RS-232 serial port 1 is a printer, a PRINT button now<br />

appears in the upper left corner of the screen. Follow these steps to print frozen<br />

graphics:<br />

1. Touch the PRINT button. The flashing message PRINTING replaces the PLOT<br />

SETUP, UNFREEZE, and PRINT buttons. You may stop printing by touching<br />

the CANCEL button.<br />

2. Once all graphics data has been sent to the printer, the PLOT SETUP,<br />

UNFREEZE, and PRINT buttons reappear.<br />

NOTE:<br />

To print graphics, RS-232 serial port 1 must be configured with PRINTER as the<br />

selected device.<br />