Dynoware Type 2825A1-2<br />

Tool Diameter<br />

Tool View<br />

Resultant Window<br />

Enter the tool diameter for the appropriate tool under test.<br />

All distances are entered in millimeters, "mm".<br />

This window shows a drawing defining the coordinate<br />

system, and tool measurement layout for the selected<br />

dynamometer.<br />

The resultant window shows channel allocation for<br />

acquired and calculated channels. This should be used as a<br />

guide when configuring graphs.<br />

6.2.1.8 Configurations Using Rotating Cutting Dyno 5237 (High Speed RCD)<br />

Selecting RCD 5237 in the amplifier selection (see section<br />

6.2.1.3 above) will enable the RCD 5237 tab, and the RS-<br />

232C Setup tab.<br />

6.2.1.8.1 RCD 5237<br />

To configure the 5237 high speed rotating cutting<br />

dynamometer, configure the following:<br />

Channel Channel number. Channels range from 1 to 2.<br />

Label<br />

Measuring Range<br />

Unit<br />

Sensitivities<br />

Range Selection<br />

Operate<br />

Reset<br />

Send Parameters<br />

Reset/Operate<br />

Assign labels for each channel. These labels appear in<br />

graphs and reports.<br />

The measuring range is automatically assigned based on<br />

the Scale and Sensitivity settings. Measuring Range [M.U.]<br />

= Full Scale Output [mV] divided by Sensitivity [mV/M.U.];<br />

(M.U. = Mechanical Unit)<br />

Assign units to each channel. These labels appear in graphs<br />

and reports.<br />

Range I, Range II and Range III – enter the sensitivity of a<br />

sensor in [mV/M.U.] for each of the calibrated ranges.<br />

Select the desired measurement range from the pull-down<br />

choice box.<br />

Transmit the command to place the amplifier into the<br />

operate mode.<br />

Transmit the command to place the amplifier into reset<br />

mode.<br />

Transmit the command to set the appropriate per channel<br />

configuration for the Range settings.<br />

Enable the automatic control (via RS-232C interface) of the<br />

Operate/Reset commands during a typical acquisition<br />

cycle. Amplifier will be automatically set to operate at the<br />

start of a cycle, and will return to reset after the acquisition<br />

has completed for any given cycle. In some cases, this<br />

automatic control of Reset/Operate is not desired and the<br />

selection box should remain unchecked.<br />