while allowing SpOz and predictive temperature<br />

determinations to be monitored and recorded<br />

(depending on Monitor model). In the Vitals (UK: All<br />

Obs) mode, the blood pressure is determined one time.<br />

Before each BP determination, the Monitor performs a test<br />

to ensure that the cuff pressure is below a specified level.<br />

The determination is delayed until this condition is met.<br />

During the delay, the BP values are displayed as zero.<br />

The Monitor senses the type of hose being used and<br />

automatically uses adult/pediatric monitoring parameters or<br />

neonatal monitoring parameters, as appropriate.<br />

Audible and visible alarms occur when a value for systolic<br />

pressure, diastolic pressure, mean arterial pressure, or pulse<br />

rate is outside the selected high or low limit.<br />

Instructions for cleaning and disinfecting BP cuffs are in<br />

Appendix F.<br />

Genera/ Warnings<br />

• The PRO Monitor will not measure blood pressure<br />

effectively on patients who are experiencing seizures<br />

or tremors.<br />

• Arrhythmias will increase the time required by the<br />

PRO Monitor to determine a blood pressure and may<br />

extend the time beyond the capabilities of the<br />

Monitor.<br />

• In <strong>Manual</strong> mode, the PRO Monitor displays the results<br />

of the last blood pressure determination for 2 minutes<br />

or until another determination is completed. If a<br />

patient’s condition changes between one<br />

determination and the next, the Monitor will not<br />

detect the change or indicate an alarm condition.<br />

• Devices that exert pressure on tissue have been<br />

associated with purpura, skin avulsion, compartmental<br />

syndrome, ischemia and/or neuropathy. To minimize<br />

these potential problems, especially when monitoring<br />

at frequent intervals or over extended periods of time,<br />

make sure the cuff is applied appropriately and<br />