General Warning<br />

• The performance of the Monitor may be degraded if it<br />

is operated outside of the environmental conditions<br />

specified in Appendix A.<br />

General Cautions<br />

• Be careful not to overextend the coiled cord of the<br />

temperature probe. Overextension can damage the<br />

probe coil connector interfaces.<br />

• Accurate oral temperatures (blue) can only be<br />

obtained by placing the probe under the tongue in the<br />

right or left sublingual pocket. Temperatures in other<br />

locations in the mouth can vary by more than 2 OF or<br />

1 OC.<br />

• Accurate rectal temperatures can only be obtained by<br />

using the red temperature probe. Red and blue<br />

temperature probes are not interchangeable.<br />

• Do not allow the tip of the predictive temperature<br />

probe to come into contact with a heat source (e.g.,<br />

hands or fingers) prior to taking a temperature<br />

determination. If this occurs, allow 5 seconds for the<br />

probe tip to cool before proceeding.<br />

• Use only IVAC* probes and probe covers. The size,<br />

shape, and thermal characteristics of the probe covers<br />

can affect the performance of the instrument.<br />

Inaccurate readings or retention problems may occur<br />

unless IVAC probes and probe covers are used.<br />

<strong>Pro</strong>cedures<br />

1. Connect the temperature probe cable to the<br />

temperature probe connector (15).<br />

2. Remove the temperature probe from the probe holster.<br />

Place a protective temperature probe cover on the<br />

probe and insert the probe appropriately.<br />

3. The determination begins automatically. Hold the<br />

temperature probe steady until the determination is<br />

complete. This takes approximately 60 seconds, during<br />

which time a pattern of lines on the temperature display<br />

(12) appears as a “chase sequence” to indicate progress.<br />