B Using the Menu System<br />
Tgt Pressure<br />
Selection of this option allows the user to set the BP target<br />
inflation pressure. The initial target pressure can be set<br />
between 100 and 250 mmHg in 5 mmHg increments. The<br />
factory default is 160 mmHg for adults and 110 for<br />
neonates. (This is indicated by “AUTO” at the end of<br />
adjustable range.) When using a neonate blood pressure cuff,<br />
if the target pressure is set to greater than 140 mmHg under<br />
the Set BP or Clinical menu, the Monitor automatically<br />
defaults to a target pressure of 110 mmHg. If the target<br />
pressure is set between 100 and 140 mmHg, then that setting<br />
is the target pressure that will be used. When the target<br />
pressure is changed, the next determination will use the new<br />
target inflation value if no systolic is available. Initial target<br />
pressure is restored to the factory default setting after<br />
power-off. The target pressure can be adjusted permanently<br />
in the Clinician menu of the Service mode (refer to “Press”<br />
in the “Using the Menu System” section).<br />
stat<br />
Selection of this option allows the user to start Stat<br />
determinations. When Stat is selected, blood pressure is<br />
determined as many times as possible in 5 minutes.<br />
Note: Alarm limits are disabled while in Stat mode.<br />
Main<br />
Selection of this button returns the user to the Main menu.<br />
Alarms Button<br />
Selection of this button displays the Alarms menu. This<br />
menu is used to adjust the violation limits for BP, Pulse Rate,<br />
and SpO,. The values and ranges for these parameters are<br />
not stored when the Monitor is turned off. The user may<br />
edit the limits, but they are restored to the default values<br />
each time the Monitor is switched on. To permanently<br />
change the alarm limits, refer to “Alarms” under “Service<br />
Button” in the “Using the Menu System” section.