120 minutes). Press the rotor to confirm the setting. After<br />
power-off, the operating mode returns to the default setting<br />
of <strong>Manual</strong>. The default setting of <strong>Manual</strong> can be overridden<br />
to return to the previous user-selected setting (Auto or<br />
<strong>Manual</strong>) by selecting Set BP (UK: BP Mode) under the<br />
Service menu.<br />
In the Auto mode, the pressure must be below 5 mmHg<br />
(neonate) or 15 mmHg (adult) for at least 30 seconds<br />
before another determination can be started. BP<br />
information will be displayed on the LED until the next<br />
determination is started. This applies to Auto mode only.<br />
Note: To cancel an Auto BP determination, select the<br />
<strong>Manual</strong> button in the Set BP menu (UK: BP Mode).<br />
Stat Mode<br />
Multiple BP readings can be taken at any time by selecting<br />
the Stat button under the Set BP button (UK: BP Mode) in<br />
the Main menu. Stat mode can also be accessed by<br />
pressing the AUTO BP key (27) and then selecting the Stat<br />
button when the Set BP menu (UK: BP Mode) appears.<br />
If a <strong>Manual</strong> determination is not in progress, a S-minute<br />
series of determinations will start. If a <strong>Manual</strong> determination<br />
is in progress, that determination will become the first in the<br />
series. A normal, uninterrupted Stat sequence will give the<br />
first set of systolic, diastolic, and mean arterial pressure<br />
values and pulse rate within 15 to 20 seconds. Selecting the<br />
Stat button during a series of Stat determinations will cancel<br />
the determination in progress and the rest of the series. BP<br />
information will be displayed on the LED until the<br />
determination has been canceled or completed. This applies<br />
to Stat mode only.<br />
The series begins with cuff inflation to a pressure above the<br />
previous systolic pressure or, if no previous systolic value is<br />
stored, to approximately 160 mmHg for adult/pediatrics.<br />
The initial target pressure selection for neonates is 110<br />
mmHg. Artifact rejection is relaxed in the Stat mode for<br />
adult/pediatric patients to allow for accelerated<br />
determinations. If a BP or Stat reading has been made<br />
previously, the first new systolic value will flash on the LED<br />
display (7) within a few seconds and will continue to flash<br />
until the end of the determination. At that point a short tone<br />
will sound and the updated systolic, diastolic, and mean<br />
30 t;