120 minutes). Press the rotor to confirm the setting. After<br />

power-off, the operating mode returns to the default setting<br />

of <strong>Manual</strong>. The default setting of <strong>Manual</strong> can be overridden<br />

to return to the previous user-selected setting (Auto or<br />

<strong>Manual</strong>) by selecting Set BP (UK: BP Mode) under the<br />

Service menu.<br />

In the Auto mode, the pressure must be below 5 mmHg<br />

(neonate) or 15 mmHg (adult) for at least 30 seconds<br />

before another determination can be started. BP<br />

information will be displayed on the LED until the next<br />

determination is started. This applies to Auto mode only.<br />

Note: To cancel an Auto BP determination, select the<br />

<strong>Manual</strong> button in the Set BP menu (UK: BP Mode).<br />

Stat Mode<br />

Multiple BP readings can be taken at any time by selecting<br />

the Stat button under the Set BP button (UK: BP Mode) in<br />

the Main menu. Stat mode can also be accessed by<br />

pressing the AUTO BP key (27) and then selecting the Stat<br />

button when the Set BP menu (UK: BP Mode) appears.<br />

If a <strong>Manual</strong> determination is not in progress, a S-minute<br />

series of determinations will start. If a <strong>Manual</strong> determination<br />

is in progress, that determination will become the first in the<br />

series. A normal, uninterrupted Stat sequence will give the<br />

first set of systolic, diastolic, and mean arterial pressure<br />

values and pulse rate within 15 to 20 seconds. Selecting the<br />

Stat button during a series of Stat determinations will cancel<br />

the determination in progress and the rest of the series. BP<br />

information will be displayed on the LED until the<br />

determination has been canceled or completed. This applies<br />

to Stat mode only.<br />

The series begins with cuff inflation to a pressure above the<br />

previous systolic pressure or, if no previous systolic value is<br />

stored, to approximately 160 mmHg for adult/pediatrics.<br />

The initial target pressure selection for neonates is 110<br />

mmHg. Artifact rejection is relaxed in the Stat mode for<br />

adult/pediatric patients to allow for accelerated<br />

determinations. If a BP or Stat reading has been made<br />

previously, the first new systolic value will flash on the LED<br />

display (7) within a few seconds and will continue to flash<br />

until the end of the determination. At that point a short tone<br />

will sound and the updated systolic, diastolic, and mean<br />

30 t;