to produce a peripheral pulse or the patient may have<br />
poor peripheral perfusion. Also, if a patient’s beat-tobeat<br />
pulse amplitude varies significantly (e.g., because<br />
of pulsus alternans, atrial fibrillation, or the use of a<br />
rapid-cycling artificial ventilator), blood pressure and<br />
pulse rate readings can be erratic, and an alternate<br />
measuring method should be used for confirmation.<br />
General Notes<br />
• A patient’s vital signs may vary dramatically during the<br />
use of cardiovascular agents such as those that raise or<br />
lower blood pressure or those that increase or decrease<br />
heart rate.<br />
9 Because treatment protocols based on the patient’s<br />
blood pressure may rely on specific values and differing<br />
measurement methods, such as auscultatory, clinicians<br />
should note a possible variance from values obtained<br />
with the PRO Monitor in planning patient care<br />
management. The PRO Monitor values are based on<br />
the oscillometric method of noninvasive blood pressure<br />
measurement and correspond to comparisons with<br />
inn-a-aortic values within ANSI /AAMI Standards for<br />
accuracy (a mean difference of + 5 mmHg, and a<br />
standard deviation of + 8 mmHg).<br />
. Several conditions may cause the BP parameter to<br />
calculate and display only the mean arterial pressure<br />
(MAP) without a systolic and diastolic reading. These<br />
conditions include very low systolic and amplitude<br />
fluctuations, so an accurate calculation for these values<br />
can’t be made (e.g., patient in shock); too small of a<br />
difference between systolic and MAP calculations in<br />
relationship to the difference between diastolic and<br />
MAP; or a leak has occurred in the PRO Monitor<br />
(1. Check all BP connections 2. Monitor may need<br />
calibration and leak testing). If only the MAP value is<br />
displayed, the systolic and diastolic will display dashes<br />
(-) and an alarm message “N99-BP FAILED” will be<br />
displayed.<br />
26