Tube focus to tabletop distance d FTD of vertical Bucky from the<br />

K i Phantom:<br />

Chest 2<br />

X ray focus: _____ mm<br />

Distance d m of dosimeter from vertical Bucky: ______ mm<br />

Anteroposterior (AP) chest<br />

thickness, t P = 225 mm.<br />

4. Dosimeter reading and calculation of incident air kerma<br />

Dosimeter reading (M 1 , M 2 , M 3 ): ___________<br />

Pressure P (kPa): ___ Temperature T ( o C): ____<br />

Mean dosimeter reading M — : _________<br />

T P<br />

kTP = Ê 273.<br />

15 + ˆ<br />

Á<br />

Ë + T<br />

˜ Ê ˆ<br />

273 15 ¯ Ë<br />

Á<br />

P ¯<br />

˜ = _____<br />

. 0<br />

0 2<br />

HVL (from Section 5 below) = ________ mm Al k Q = ________<br />

Calculated value of air kerma, K(d) = M — N K,Q0 k Q k TP = ____________________ mGy<br />

Calculated value of incident air kerma,<br />

2<br />

Ê d d<br />

K K d<br />

FTD - m<br />

ˆ<br />

_______<br />

i = ( ) Á<br />

mGy.<br />

Ë d FTD -<br />

˜ =<br />

225¯<br />

5. Determination of HVL<br />

Dosimeter readings should be obtained for filter thicknesses that bracket the HVL. The first and<br />

last readings, M 01 and M 02 , are made at zero filter thickness.<br />

Filter thickness<br />

(mm Al)<br />

Dosimeter reading, M<br />

(mGy)<br />

Average dosimeter reading, M — , at zero<br />

thickness<br />

0.00 (M 01 + M 02 )/2 = _________ mGy<br />

0.00 Interpolated HVL: ______ mm Al<br />

2<br />

For dosimeters with a semiconductor detector, k TP = 1.<br />

179