DISPLAY PANEL MENU KEY<br />

USER'S MANUAL<br />

Display Panel Menu Key Descriptions (Cont.)<br />

System Menu 7: Restoring Factory Default Settings<br />

Restore factory defaults restores the original factory set voltages and<br />

waveforms in the 10 waveform storage locations. All the current<br />

settings will be changed back to the Factory Default Settings. See<br />

Appendix D for a list of Factory Default Settings.<br />

Fig. 2.19<br />

This message will be displayed before the default settings are restored<br />

and the user must press the Up arrow key to confirm restoration of t<br />

those settings (See Fig. 2.20). Pressing any other key will abort this<br />

operation and return to the status display.<br />

Fig. 2.20<br />

System Menu 8: Event History<br />

The <strong>LR</strong>-<strong>24</strong> records all errors and unusual events that occur during<br />

use. This menu option allows a technician to review the status codes<br />

stored during field use of the <strong>LR</strong>-<strong>24</strong> (See Fig. 2.21).<br />

This history file is an aid to troubleshooting if the <strong>LR</strong>-<strong>24</strong> develops<br />

a problem. There is a table of error codes in Appendix B of this<br />

manual. To exit from this mode and return to the status screen. Press<br />

the enter key (See Fig. 2.22).<br />

Fig. 2.21<br />

Fig. 2.22<br />

www.smith-root.com<br />