Table 33: Nominal <strong>Pulse</strong> Generator LongevityAmplitudeImpedance(Ohms)<strong>Estella</strong> <strong>Pulse</strong> <strong>Generators</strong> Technical Manual 119Pacing10% 50% 100%2.5 V 500DR(-T)SR(-T)DR(-T)SR(-T)DR(-T)SR(-T)12.1 15.0 10.7 14.1 8.2 11.9The mean* expected time intervals from ERI to EOS at standardprogram for <strong>Estella</strong> pulse generators is 6 months. All service intervals,including the above-cited nominal pulse generator longevity, arebased on considerations that consider the battery discharge behaviorand the hybrid circuit properties including current consumption andreplacement indicator. The statistical calculations are based on a basicrate of 60 ppm, pulse width of 0.4 ms, shelf life of 1.5 years, and datasupplied by the battery manufacturer.* 50% of all pacemakers reach or exceed the given value