126 <strong>Estella</strong> <strong>Pulse</strong> <strong>Generators</strong> Technical ManualNight rateOff, 30...(1)...60...(1)...88...(2)...122...(3)...140...(5)...200 ppmRate HysteresisOff; -5...(5)...-90 bpmRepetitive HysteresisOff; 1…(1)…15Scan HysteresisOff: 1…(1)…15Upper Rate90…(10)…130…(10)…200 ppmV pSuppressionOff; OnUTR ResponseWRL (automatic selection)Rate Limitation*,†,‡190…220 ppmDynamic AV Delay (Dual chamber only)Off; low; medium; high, (individual, or fixed), I-OptAV Delay Values (Dual chamber only)15…(5)…180…(5)…350 ms(Programmable in 5 ranges)AV Delay Hysteresis (Dual chamber only)Off; low; medium; high; negative, IOPT* The corresponding intervals t correlate with the rates f by the formula t = 60.000/ f (t in ms, f in ppm).† In the event of electronic defect.‡ Rate Limitation changes as the Pacemaker approaches End of Service. TheRate Limitation is nominally 190 ppm at Beginning of Service (BOS) and can reach220 ppm at End of Service (EOS) due to battery depletion.