The AAI mode is conditionally indicated if<strong>Estella</strong> <strong>Pulse</strong> <strong>Generators</strong> Technical Manual 141• the hemodynamics of patients with bradycardia and symptomaticallyreduced cardiac output can be improved by raising the heart rate,given that adequate AV conduction has been established by anappropriate diagnostic examination.The AAI mode is contraindicated for• previously established AV conduction delay or AV block or ifdiminishing AV conduction has been determined by appropriatetests, and/or• inadequate intra-atrial complexes that do not permit safe sensing.17.1.5 Other ModesIn addition to the ACC/AHA guidelines, the modes listed above mayhave further indications due to medical/technical complications such aselectromagnetic interference, sensing defects, fracture of the lead(s),detection of myopotentials, muscle stimulation, etc. The same appliesto the asynchronous DOO(R), AOO(R) and VOO(R) pacing modes derivedfrom the above by restricting the sensing functions [SOO(R) modeavailable with <strong>Estella</strong> SR models]. The triggered DDT, AAT and VVTpacing modes and the VDI and OFF modes are indicated for diagnosticpurposes to assess intrinsic cardiac activity. Use of the OFF mode iscontraindicated in pacemaker dependent patients.