10.07.2015 Views

Estella Pulse Generators - BIOTRONIK USA - News

Estella Pulse Generators - BIOTRONIK USA - News

Estella Pulse Generators - BIOTRONIK USA - News

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Estella</strong> <strong>Pulse</strong> <strong>Generators</strong> Technical Manual 89Transmitted Device SettingsThe primary programmed parameters for the following are sent in thedata package:• Leads—(e.g., Pacing Output, Configuration)• Brady—(e.g., Basic Rate, UTR, AV-Delays, RV Sensitivity)• IOPT—(ON/OFF)• HM Settings—(e.g., ON/OFF, transmission time (daily), IEGMtransmissions ON/OFF, periodic IEGM, ongoing atrial episode,statistics, holter)• Miscellaneous other information7.17.5.1 IEGM Online HDsThe <strong>Estella</strong> provides the ability to transmit periodic IEGM Online HD(IEGM and marker data) from the periodic follow-ups as an addition tothe current messages.An IEGM with up to 2 channels (RV and/or RA) are sent in one message,depending on the number of IEGM channels programmed in the Holterconfiguration.The following markers are also transmitted: A S(including A rs), A S(PMT),A P, V S(including V rs), V P(atrial and refractory sensed events includedwith sensed events).The <strong>Estella</strong> includes a programmable parameter to disable or enablethe IEGM transmission.