<strong>Estella</strong> <strong>Pulse</strong> <strong>Generators</strong> Technical Manual 154.6.3 Hospital and Medical EnvironmentsElectrosurgical Cautery—Electrosurgical cautery could induceventricular arrhythmias and/or fibrillation, or may cause asynchronousor inhibited pulse generator operation. If use of electrocautery isnecessary, the current path (ground plate) should be kept as far awayfrom the pulse generator and leads as possible.Lithotripsy—Lithotripsy may damage the pulse generator. If lithotripsymust be used, do not focus the beam near the pulse generator.External Defibrillation—External defibrillation may damage the pulsegenerator. Attempt to minimize current flowing through the pulsegenerator and lead system by following the precautions.High Radiation Sources—High radiation sources such as cobalt 60 orgamma radiation should not be directed at the pulse generator. If apatient requires radiation therapy in the vicinity of the pulse generator,place lead shielding over the device to prevent radiation damage.4.7 <strong>Pulse</strong> Generator Explant and DisposalDevice Incineration—Never incinerate a pulse generator. Be sure thepulse generator is explanted before a patient who has died is cremated(see Section 14).Explanted Devices—Return all explanted devices to <strong>BIOTRONIK</strong>.