10.07.2015 Views

Estella Pulse Generators - BIOTRONIK USA - News

Estella Pulse Generators - BIOTRONIK USA - News

Estella Pulse Generators - BIOTRONIK USA - News

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Estella</strong> <strong>Pulse</strong> <strong>Generators</strong> Technical Manual 33Time from Event Onset toEvaluation of First Event/PatientGroup 1N=972Group 2N=471p valueVF1.010.5 +/- 22.20145236 (24.3%)SVT2.016.6 ± 27.4010894 (9.7%)MedianMean ± SD (days)MinMax# of patients with events35.545.0 +/- 47.0028792 (19.5%)p < 0.001p < 0.001MedianMean ± SD (days)MinMax# of patients with events39.042.1 ± 35.6015735 (7.4%)p < 0.001p < 0.001Figure 4: Median Time from Onset to Evaluation of Arrhythmic EventsAnalysisThe mean time from onset to evaluation of first AF, VT, and VF events inGroup 2 is greater than the mean time from onset to evaluation of firstAF, VT, or VF events in Group 1. A rejection of the null hypothesis forAF, VT and VF event types indicates that the mean time from onset toevaluation of the first AF, VT and VF events in Group 1 is significantly lessthan the mean time from onset to evaluation of the first AF, VT and VFevents in Group 2. P-values are =0.005,