7.5 Loadcell calibrationWhen using a <strong>MC</strong>-<strong>Balance</strong> for the first time do an initial load cell calibration as follows:• The unit must be mounted horizontally (water leveled)• Avoid vibrations during the load cell calibration. This will influence the calibration.• Do not touch the unit during load cell calibration.• When using a slide the whole unit has to be slid in against the neckpiece and fixed.• Be sure that the <strong>Balance</strong> load cell is connected to the controller• Go to the LOAD CELL • In this menu it is possible to calibrate the load cell of <strong>MC</strong>-<strong>Balance</strong>(500gr. Calibration weight required)• Select YES to start the load cell calibrationLOADCELL CALIBRATION[***][***] = ProgressStatusCalibration busy.Please wait…………..• After a few minutes the following screen appears:LOADCELL CALIBRATION[***]StatusPlace weight, pressSTART when ready.Place calibration weight (500 gr.) on the hopperand press the START button• After approx. 1 minute the load cell calibration is ready, press the MENU button to continue.• To check if the load cell calibration was OK, go to the WEIGHT CHECK WEIGHT CHECKWeight: gr. >