5.2 After placement of purchase order, vendor shall submit certified drawings and<br />

<strong>specification</strong> sheets <strong>for</strong> each chromatograph which shall include the following as a<br />

minimum:a)<br />

Overall dimensions in mm <strong>for</strong> each sub system and its weight.<br />

b) Detailed interconnection diagram <strong>for</strong> wiring, process piping and tubing.<br />

c) System wiring diagram including chromatograph, programmer and<br />

d)<br />

communication interface.<br />

Installation details.<br />

e) Detailed operating and maintenance manuals <strong>for</strong> each sub system.<br />

f) Cabling details including shielding/grounding requirements.<br />

g) Factory testing and inspection requirement.<br />

5.3 Vendor shall provide certificate <strong>for</strong> all the tests indicated in para 4.0 of this<br />

<strong>specification</strong>. In addition, vendor shall provide the 'Manufacturer's Certificate of<br />

Con<strong>for</strong>mity to Purchaser's Specifications' as per clause 2.2 of DIN-50049.<br />

5.4 The final as-built documentation shall be supplied by the vendor properly bound in A4<br />

size i.e. 216 x 279 mm sheets. All drawings and sketches shall be in multiple of A4 size<br />

but shall be folded to this size.<br />

6.0 DESIGN AND CONSTRUCTION REQUIREMENTS<br />

6.1 GENERAL DEFINITIONS<br />

The following components of the gas chromatograph system are defined as below:<br />

a) Chromatograph:- A stand alone, unit architecture based microprocessor<br />

controlled instrument consisting of chromatograph proper, sample conditioning<br />

system and programmer which is capable of working and generating hardwired<br />

and serial outputs without the aid of any external device.<br />

b) Programmer:- A microprocessor based device which is a part of chromatograph<br />

and controls the overall functioning of the system and also provides hardwired<br />

analogy and serial outputs.<br />

c) Communication Interface:- A high level microprocessor based device that can<br />

"talk" to a number of programmers and does the job of data collection,<br />

concentration and providing a single serial interface between chromatograph<br />

system and purchaser's DCS or host computer system.<br />

STANDARD SPECIFICATION SPECIFICATION REV-0<br />

<strong>GAIL</strong> (INDIA) LTD<br />

FOR<br />

NEW DELHI<br />

GAS CHROMATOGRAPH S-08-02-025 Page 7 of 17<br />

Copyright <strong>GAIL</strong> – All rights reserved