Gas Gr. IIA & IIB Temp T3. All analyzer components located in hazardous areas must<br />

meet the electrical classification of that area.<br />

6.1.13 Analyzer shall be selected <strong>for</strong> high accuracy and repeatability as the primary criteria.<br />

Analysis speed and cycle time is secondary <strong>for</strong> custody transfer purposes.<br />

6.1.14 Analyzer and sample system shall be suitable <strong>for</strong> outdoor installation in the field.<br />

Minimum of IP66 protection shall be provided.<br />

6.1.15 A sunshade and a three-side weather shelter shall be provided to protect the Analyzer<br />

and sample system.<br />

6.1.16 Power supply <strong>for</strong> the gas chromatograph system shall be 24 VDC / 230 Volts ac from<br />

the purchasers electrical supply system. VENDOR shall provide the power requirements<br />

of the system.<br />

6.2 SAMPLE SYSTEM<br />

6.2.1 Vendor shall provide a fabricated assembly of sample conditioning system having facility to<br />

measure sample flow to the chromatograph proper.<br />

6.2.2 Sampling system design should be such that the sample drawn <strong>for</strong> analysis is truly<br />

representative of the process stream. The unit shall be supplied with all equipment necessary to<br />

condition each stream <strong>for</strong> injection into the chromatograph. The system must include, but not<br />

limited to, all Regulators, vaporizers, dryers, filters, rotameter, solenoid valves required <strong>for</strong><br />

proper analysis. All Components of sampling system shall be rated <strong>for</strong> the mechanical design<br />

pressure and temperature conditions indicated in purchaser's data sheet.<br />

6.2.3 Transportation time <strong>for</strong> the sample from the process tapping point upto the sampling system<br />

shall not exceed 30% of the cycle time specified <strong>for</strong> the chromatograph.<br />

6.2.4 The sampling system shall contain at least one wire mesh strainer of US <strong>standard</strong> sieve mesh<br />

size 100 or finer. Where purchaser's data sheets indicate greater possibilities of plugging,<br />

vendor shall provide more filters/separators as required.<br />

6.2.5 Where purchaser's data sheets indicate possibilities of polymer <strong>for</strong>mation on presence of excess<br />

polymers, vendor shall offer suitable design features to minimize plugging because of polymer<br />

residue.<br />

6.2.6 Wherever the process stream pressure is low, vendor shall provide/ advise the purchaser on<br />

use of aspirators or sample pumps. Vendor shall also indicate the utility requirements (e.g water<br />

or air) <strong>for</strong> such devices.<br />

6.2.7 Sampling system shall include provisions <strong>for</strong> connection of calibration <strong>standard</strong>.<br />

6.2.8 Sampling system shall include block valves on all process sampling lines.<br />

<strong>GAIL</strong> (INDIA) LTD<br />

NEW DELHI<br />

STANDARD SPECIFICATION<br />

FOR<br />

GAS CHROMATOGRAPH<br />

SPECIFICATION<br />

S-08-02-025<br />

REV-0<br />

Page 9 of 17<br />