Notes<br />

Calculation<br />

Copyright 2009, IMPAC Medical Systems<br />

Appendix B: Dose Calculation<br />

1. The Dose Site Dose is calculated for each historical treatment record:<br />

DSD = TMU * FDC * DSC.<br />

2. The Dose Site Doses from dose history records for the same treatment field and<br />

within the same treatment session are added together.<br />

3. The same field/same session DSD is rounded to the nearest cGray (or Gray<br />

depending on the dose unit configured at your facility in the Dosimetry<br />

Configuration dialog box).<br />

4. For each field/session, the rounded dose is added to the running total to give the<br />

Total Dose.<br />

The Total Dose value calculated above is the Total Dose that appears in the<br />

SEQUENCER table, eCHART Check, Dose Site Summary window and report,<br />

and in the Treatment Chart.<br />

The Daily Treatment Summary still shows field dose rounded<br />

to the nearest cGray per historical treatment, where Field<br />

Dose = TMU * FDC.