Section 1 Introduction<br />

tip of the needle. This is generally due to a misalignment<br />

between the needle and the bottle. The small leak does not affect<br />

sample integrity, as it occurs outside the bottle before the valve is<br />

closed. If the leak condition persists or sample is being sprayed<br />

out of this connection, consult Teledyne <strong>Isco</strong> Customer Service. A<br />

leak around the valve stem on the top of the bottle indicates a<br />

worn valve. When this occurs, the valve body must be replaced.<br />

One bottle does not have a valve on top. This is the blank, and<br />

you will also notice that the bottle rack looks different behind<br />

this bottle. There is no slot in the rack. This bottle marks the<br />

home position for the bottle rack. When you reload the bottles<br />

after sampling always put the blank bottle back in this position,<br />

or the sampler will not work properly.<br />

1.3.3 Cleaning the Bottles<br />

(the First Time)<br />

Teledyne <strong>Isco</strong> assembles the racks with new clean bottles, as<br />

shipped from the bottle manufacturer. Normally, no further<br />

cleaning is necessary before first use of the bottles. However,<br />

these bottles and valves are not certified clean. If you wish to<br />

clean the bottles before using them the first time (for example if<br />

you are sampling for extremely low concentration volatile<br />

organic compounds), you may want to clean them. Follow EPA<br />

guidelines for washing and drying the bottles. After you have<br />

used the bottles for samples, always wash and dry them<br />

according to EPA guidelines.<br />

CAUTION<br />

Always make sure the rack is in the "home" position before<br />

installing or removing the bottle rack. This places the blank bottle<br />

under the fill needle. If the rack is not in "home" position,<br />

press Rack Reset to return the rack before removing or installing<br />

the bottle rack. If you do this, the rack will always be in the<br />

right position when you reload it.<br />

1.3.4 Installing the Bottle<br />

Rack<br />

1.3.5 Attaching the Sample<br />

Tubing<br />

1.3.6 Attaching the Drain<br />

Tube<br />

Notice the D-shaped steel shaft in the center of the triangle<br />

formed by the three knobs. This is the bottle rack shaft. Place the<br />

bottle rack over this shaft, rotating it slowly until the D-shape of<br />

the shaft lines up with the same D-shape on the bottom of the<br />

bottle rack. The rack will easily settle into place. Secure the rack<br />

in place by tightening the knurled knob and threaded shaft into<br />

the threaded hole on the D-shaped shaft on the controller base.<br />

The rack is now ready to take samples.<br />

The <strong>6100CR</strong> uses Teflon 1 bonded polyethylene tubing to connect<br />

to the source and to the filter. Connection of the tubing is with<br />

1 /4" tubing fittings.<br />

Teledyne <strong>Isco</strong> supplies a 20 foot length of 1 /4" OD vinyl tubing to<br />

drain the sampler overfill and sample line bypass drain into any<br />

convenient place that gets the water out of the way, and cannot<br />

flow back toward the <strong>6100CR</strong>. There is a restrictor in the end of<br />

the drain line. This must remain in place.<br />