<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />
Section 1 Introduction<br />
tip of the needle. This is generally due to a misalignment<br />
between the needle and the bottle. The small leak does not affect<br />
sample integrity, as it occurs outside the bottle before the valve is<br />
closed. If the leak condition persists or sample is being sprayed<br />
out of this connection, consult Teledyne <strong>Isco</strong> Customer Service. A<br />
leak around the valve stem on the top of the bottle indicates a<br />
worn valve. When this occurs, the valve body must be replaced.<br />
One bottle does not have a valve on top. This is the blank, and<br />
you will also notice that the bottle rack looks different behind<br />
this bottle. There is no slot in the rack. This bottle marks the<br />
home position for the bottle rack. When you reload the bottles<br />
after sampling always put the blank bottle back in this position,<br />
or the sampler will not work properly.<br />
1.3.3 Cleaning the Bottles<br />
(the First Time)<br />
Teledyne <strong>Isco</strong> assembles the racks with new clean bottles, as<br />
shipped from the bottle manufacturer. Normally, no further<br />
cleaning is necessary before first use of the bottles. However,<br />
these bottles and valves are not certified clean. If you wish to<br />
clean the bottles before using them the first time (for example if<br />
you are sampling for extremely low concentration volatile<br />
organic compounds), you may want to clean them. Follow EPA<br />
guidelines for washing and drying the bottles. After you have<br />
used the bottles for samples, always wash and dry them<br />
according to EPA guidelines.<br />
CAUTION<br />
Always make sure the rack is in the "home" position before<br />
installing or removing the bottle rack. This places the blank bottle<br />
under the fill needle. If the rack is not in "home" position,<br />
press Rack Reset to return the rack before removing or installing<br />
the bottle rack. If you do this, the rack will always be in the<br />
right position when you reload it.<br />
1.3.4 Installing the Bottle<br />
Rack<br />
1.3.5 Attaching the Sample<br />
Tubing<br />
1.3.6 Attaching the Drain<br />
Tube<br />
Notice the D-shaped steel shaft in the center of the triangle<br />
formed by the three knobs. This is the bottle rack shaft. Place the<br />
bottle rack over this shaft, rotating it slowly until the D-shape of<br />
the shaft lines up with the same D-shape on the bottom of the<br />
bottle rack. The rack will easily settle into place. Secure the rack<br />
in place by tightening the knurled knob and threaded shaft into<br />
the threaded hole on the D-shaped shaft on the controller base.<br />
The rack is now ready to take samples.<br />
The <strong>6100CR</strong> uses Teflon 1 bonded polyethylene tubing to connect<br />
to the source and to the filter. Connection of the tubing is with<br />
1 /4" tubing fittings.<br />
Teledyne <strong>Isco</strong> supplies a 20 foot length of 1 /4" OD vinyl tubing to<br />
drain the sampler overfill and sample line bypass drain into any<br />
convenient place that gets the water out of the way, and cannot<br />
flow back toward the <strong>6100CR</strong>. There is a restrictor in the end of<br />
the drain line. This must remain in place.<br />
1. Teflon ® , registered trademark of du Pont de Nemours, Inc.<br />
1-5