Section 2 Programming<br />

Note<br />

You must have a special plug that loops the transmit and<br />

receive signals together to run the “TEST PRINTER” and<br />

“FLOW METER” diagnostic.<br />

16. This option causes the sampler to test the flow meter serial<br />

port by sending and receiving data. As for the TEST<br />

PRINTER PORT diagnostic above, you must have a special<br />

plug to connect the appropriate pins.<br />

TEST FLOW METER PORT<br />

YES NO<br />

17. This option tests the mechanical operations of the sampler<br />

by moving the tower up and down, operating the valve<br />

driver, and rotating the bottle rack one full revolution.<br />

TEST OPERATION<br />

YES NO<br />

18. The SELECT DISABLE MODE menu refers to the treatment<br />

of sampling events that occur during the time the<br />

sampler is disabled, or shut down. If time-paced, the sampler<br />

will have three disable choices.<br />

SELECT DISABLE MODE<br />

ERROR SKIP DELAY<br />

Note<br />

The Select Disable Mode choice only applies if the <strong>6100CR</strong> is<br />

programmed for time-paced sampling. It determines operation<br />

if samples are missed either because of the unit being disabled,<br />

or because of power failure. It does not cover samples<br />

missed because the unit has been shut down. For units programmed<br />

to take flow-paced samples, the sampler will take a<br />

sample on enable and reset the flow pulse counter.<br />

• If you select SKIP, the sampler will maintain the time<br />

intervals as set up at start time. The log will show the<br />

event as an error, and the sampler will skip the bottle(s)<br />

allocated to that time and advance to the next bottle<br />

position so that following samples will be placed in the<br />

scheduled bottles.<br />

• Selecting ERROR means that the sampler will log an<br />

error. The time interval will reset. If it is programmed for<br />

sample intervals of one hour and the sampler is disabled<br />

2-8