Section 2 Programming<br />
Note<br />
You must have a special plug that loops the transmit and<br />
receive signals together to run the “TEST PRINTER” and<br />
“FLOW METER” diagnostic.<br />
16. This option causes the sampler to test the flow meter serial<br />
port by sending and receiving data. As for the TEST<br />
PRINTER PORT diagnostic above, you must have a special<br />
plug to connect the appropriate pins.<br />
TEST FLOW METER PORT<br />
YES NO<br />
17. This option tests the mechanical operations of the sampler<br />
by moving the tower up and down, operating the valve<br />
driver, and rotating the bottle rack one full revolution.<br />
TEST OPERATION<br />
YES NO<br />
18. The SELECT DISABLE MODE menu refers to the treatment<br />
of sampling events that occur during the time the<br />
sampler is disabled, or shut down. If time-paced, the sampler<br />
will have three disable choices.<br />
SELECT DISABLE MODE<br />
ERROR SKIP DELAY<br />
Note<br />
The Select Disable Mode choice only applies if the <strong>6100CR</strong> is<br />
programmed for time-paced sampling. It determines operation<br />
if samples are missed either because of the unit being disabled,<br />
or because of power failure. It does not cover samples<br />
missed because the unit has been shut down. For units programmed<br />
to take flow-paced samples, the sampler will take a<br />
sample on enable and reset the flow pulse counter.<br />
• If you select SKIP, the sampler will maintain the time<br />
intervals as set up at start time. The log will show the<br />
event as an error, and the sampler will skip the bottle(s)<br />
allocated to that time and advance to the next bottle<br />
position so that following samples will be placed in the<br />
scheduled bottles.<br />
• Selecting ERROR means that the sampler will log an<br />
error. The time interval will reset. If it is programmed for<br />
sample intervals of one hour and the sampler is disabled<br />
2-8