<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 2 Programming<br />

should always discharge into a normal, non-pressurized<br />

open drain.<br />

2. If you select YES, the following will appear:<br />

PUT DISCHARGE TUBE IN DRAIN<br />

Alternating with:<br />

Make sure that the sampler drain line is in a suitable<br />

drain and press the Manual Purge key. This will start liquid<br />

flow through the sample cooler and discharge through<br />

the drain line.<br />

Followed by:<br />

THEN,<br />

PRESS MANUAL PURGE<br />

Observe the discharge line. After the air is purged from the<br />

sample cooler, you should only observe a few bubbles in the<br />

sample line. When the air is purged, push Clear/Exit, and<br />

the flow will stop.<br />

Followed by:<br />

WHEN NO BUBBLES<br />

PRESS CLEAR/EXIT<br />

PUT DISCHARGE TUBE<br />

IN GRAD CYLINDER<br />

Alternating with:<br />

THEN,<br />

PRESS MANUAL SAMPLE<br />

Place the discharge tube into a beaker or graduated cylinder.<br />

Place the beaker at the same level as the discharge<br />

tube will be in normal operation. Then press the Manual<br />

Sample key to start the 50 second flow for calibration. The<br />

display will show:<br />

COLLECTING 100 ML<br />

TEST VOLUME XX<br />

2-3