Section 3 Installation of the <strong>Sampler</strong><br />
CAUTION<br />
Proper orientation of the filter at installation and proper reassembly<br />
after cleaning are essential for correct operation. Correct<br />
assembly and orientation ensure that particulate material<br />
trapped inside the filter will remain outside the filter screen<br />
and be easily removable. Reverse assembly of the filter will<br />
cause this material to fill the inside of the screen and plug it<br />
completely. This material is very difficult to remove once it is<br />
firmly packed inside the screen element.<br />
Do not under any circumstances operate the filter without the<br />
screen element in place. If it becomes clogged and cannot be<br />
cleared, replace it, but always make sure there is a screen in<br />
place. Operation without the screen will cause clogging in the<br />
fill needle apparatus that cannot be removed, and will require<br />
replacement of the needle, and may cause further harm to the<br />
sampler’s mechanism.<br />
Note<br />
The recommended maintenance cycle for cleaning the primary<br />
(large) filter is once per filled bottle rack inside the sampler (24<br />
bottles.)<br />
Flow<br />
Arrow<br />
Arrow<br />
Figure 3-2 Detail of Filter Installation<br />
3-3