07.03.2014 Views

6100CR Refrigerated Sampler - Isco

6100CR Refrigerated Sampler - Isco

6100CR Refrigerated Sampler - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 2 Programming<br />

2.8 Error Messages The possible ERROR types are:<br />

• RACK JAM<br />

• TOWER JAM<br />

• RACK FULL<br />

• VALVE JAM<br />

• POWER FAILURE<br />

If the RACK JAM message appears, try to clear the problem by<br />

pressing the Rack Reset key. If TOWER JAM appears, call<br />

Teledyne <strong>Isco</strong> Customer Service. RACK FULL means the<br />

sampler has filled all 24 bottles, and you will need to replace the<br />

bottle rack with a new one. VALVE JAM indicates the sampler<br />

was unable to open the valve on the sample bottle. Check the<br />

bottle in question and make changes as necessary.<br />

POWER FAILURE means that the power was disconnected<br />

during the sampling routine.<br />

If there is some problem and the unit is unable to take a sample<br />

two times in a row because of the same error, an ERROR<br />

message will appear on the display:<br />

*SAMPLING ERROR*<br />

(error type)<br />

2.9 The Display Status Key<br />

and the History Log<br />

1. When you press the Display Status key, the sampler will<br />

show you a log of events it has recorded in memory. The<br />

first display will ask you to choose from the following:<br />

DISPLAY STARTING AT<br />

BEGINNING LAST SAMP<br />

2. If you select LAST SAMP, the display will only show you<br />

the record for the last program the <strong>6100CR</strong> has run.<br />

3. If you choose BEGINNING, the log will go as far back as<br />

the internal memory allows. It will either go to the first<br />

entry made or to the oldest entry in memory. (When memory<br />

allocated to the log is full, the log will begin overwriting<br />

the oldest entries with new ones.) If you choose BEGIN-<br />

NING and the entries are too long or not what you want,<br />

you can leave the log by pressing the Clear/Exit key.<br />

Note<br />

In general, LAST SAMP is the better choice for viewing the log.<br />

Do not select BEGINNING unless you need to look over the<br />

entire log. There may be a large number of entries. If the sampler<br />

has been running for some time, the log can have as<br />

many as 1,000 entries, going back several months. Trying to<br />

make sense of all this when you can only see two lines at a<br />

time could be difficult. You can use the Clear/Exit key to leave<br />

the log at any point.<br />

2-16

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!