6100CR Refrigerated Sampler - Isco
6100CR Refrigerated Sampler - Isco
6100CR Refrigerated Sampler - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />
Section 2 Programming<br />
2.8 Error Messages The possible ERROR types are:<br />
• RACK JAM<br />
• TOWER JAM<br />
• RACK FULL<br />
• VALVE JAM<br />
• POWER FAILURE<br />
If the RACK JAM message appears, try to clear the problem by<br />
pressing the Rack Reset key. If TOWER JAM appears, call<br />
Teledyne <strong>Isco</strong> Customer Service. RACK FULL means the<br />
sampler has filled all 24 bottles, and you will need to replace the<br />
bottle rack with a new one. VALVE JAM indicates the sampler<br />
was unable to open the valve on the sample bottle. Check the<br />
bottle in question and make changes as necessary.<br />
POWER FAILURE means that the power was disconnected<br />
during the sampling routine.<br />
If there is some problem and the unit is unable to take a sample<br />
two times in a row because of the same error, an ERROR<br />
message will appear on the display:<br />
*SAMPLING ERROR*<br />
(error type)<br />
2.9 The Display Status Key<br />
and the History Log<br />
1. When you press the Display Status key, the sampler will<br />
show you a log of events it has recorded in memory. The<br />
first display will ask you to choose from the following:<br />
DISPLAY STARTING AT<br />
BEGINNING LAST SAMP<br />
2. If you select LAST SAMP, the display will only show you<br />
the record for the last program the <strong>6100CR</strong> has run.<br />
3. If you choose BEGINNING, the log will go as far back as<br />
the internal memory allows. It will either go to the first<br />
entry made or to the oldest entry in memory. (When memory<br />
allocated to the log is full, the log will begin overwriting<br />
the oldest entries with new ones.) If you choose BEGIN-<br />
NING and the entries are too long or not what you want,<br />
you can leave the log by pressing the Clear/Exit key.<br />
Note<br />
In general, LAST SAMP is the better choice for viewing the log.<br />
Do not select BEGINNING unless you need to look over the<br />
entire log. There may be a large number of entries. If the sampler<br />
has been running for some time, the log can have as<br />
many as 1,000 entries, going back several months. Trying to<br />
make sense of all this when you can only see two lines at a<br />
time could be difficult. You can use the Clear/Exit key to leave<br />
the log at any point.<br />
2-16