<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 1 Introduction<br />

General<br />

Table 1-1 <strong>Isco</strong> <strong>6100CR</strong> Cluster Rule <strong>Sampler</strong> Specifications<br />

Height 47 in. 117 cm<br />

Width 26 in. 66 cm<br />

Depth 26 in. 66 cm<br />

Weight (Dry) 190 lb. 86 kg<br />

Refrigerator Body Material<br />

Power Requirements<br />

Fiberglass-reinforced plastic with UV-resistant gel coat.<br />

120 Volts, 60 Hz<br />

Operational Temperature Range 32° to 120° F 0° to 50° C<br />

Controller<br />

Enclosure Rating<br />

Interfacing<br />

Diagnostics<br />

Clock Accuracy<br />

Sample Collection<br />

Sample (Vial) Volume Vial Capacity<br />

Rinse Cycles<br />

Sample Frequency<br />

Time Mode<br />

Flow Pacing<br />

Sample Pacing<br />

Multiplexing<br />

Sample Line<br />

Self-certified NEMA 4X and IP67<br />

6-pin connector; data output at 2400 baud in ASCII RS-232 format allows<br />

transfer of Program Settings Report, Sampling Results Report, and history<br />

log. Compatible with <strong>Isco</strong> Samplink software.<br />

Tests RAM, ROM, all mechanical drives, display, flow meter interface, and<br />

printer interface.<br />

1 minute per month typical<br />

40 ml (fixed volume)<br />

24 vials with Teflon valves<br />

1 vial reserved as blank for procedural use.<br />

All wetted surfaces automatically rinsed prior to each sample collection<br />

15 minutes to 99 hours 59 minutes, in one minute increments.<br />

Selectable, 1-9,999 pulses in single pulse intervals.<br />

Time or flow. Flow mode paced by pulses from an external flow meter.<br />

Multiple vial per sample event only. As many as 24 vials can be filled by one<br />

sample initiation.<br />

TPE Teflon lined tubing<br />

Sample Source Pressure Minimum 15 psi Maximum 120 psi<br />

Sample Source Temp.<br />

Sample Filter<br />

Sample Pre-cooling<br />

85°C Maximum<br />

Stainless steel in-line cartridge style. 125 ml capacity<br />

Sample pre-cooled to 25°C (for 85° sample source, 40° ambient temperature,<br />

and 2-hour sampling interval ).<br />