<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 1 Introduction<br />

sampler previously detected a fault in the tower, it will test the<br />

tower before it rotates the rack. The same will occur if the<br />

sampler previously detected a fault in the valve actuator, and the<br />

valve actuator must be returned to the normal position if it is<br />

not. Also, if the needle is not fully raised, the sampler will raise it<br />

before rotating the rack. The sampler raises the needle and<br />

returns the valve actuator to the normal position so they will not<br />

interfere with the rotation of the rack. Rack Reset will not have<br />

an effect when the sampler is in the running state.<br />

1.4.11 Manual Purge Pressing the Manual Purge key when the sampler is in the<br />

standby state will bring up the following screen:<br />

MANUAL PURGE<br />

STRAINER PRE-COOLER<br />

Selecting strainer will cause the sampler to open the flow valve<br />

so liquid will flow from the sample source through the strainer<br />

and valve. The liquid will be diverted by the valve and will not go<br />

through the sample pre-cooler or the actual bottle-filling mechanism,<br />

but will flow out through the bypass drain line. The flow<br />

can be stopped by pressing the ON/OFF key, or the Clear/Exit<br />

key. The Purge will also be stopped after a fifteen minute<br />

time-out. This function is useful for checking out the operation of<br />

the system, as well as purging air from the system after the<br />

strainer housing has been opened for cleaning.<br />

Similarly, selecting PRE-COOLER will cause the sampler to open<br />

the valve so the liquid will flow through the strainer, pre-cooler<br />

and the needle mechanism that fills the sample collection bottles.<br />

None will flow through the bypass. It exits the system through<br />

the sample overflow drain line. This selection does not cause the<br />

sampler to open a sample vial or put the sampling needle mechanism<br />

into a sample vial.<br />

Therefore, no liquid is introduced into a vial and no vials are contaminated<br />

by this operation. This operation is useful to verify<br />

that the system is functional, and that the valves open properly<br />

so the liquid can flow through the system. The operation is also<br />

useful to purge air from the pre-cooler upon initial installation<br />

and if it is drained during subsequent maintenance operations.<br />

The Manual Purge selections can also be useful if the sampler is<br />

not to be used for some period of time and it is desired to flush<br />

the system with clean water. If the inlet line is disconnected from<br />

the sample source and connected to clean water, these selections<br />

will make the sampler open the valves so clean water can flow<br />

through the system.<br />

1.4.12 Manual Sample This key allows you to take a sample any time the <strong>6100CR</strong> is in<br />

Standby, but it will not do manual sample within a running<br />

program. When you press this key, the unit will place a sample in<br />

the next bottle.<br />