<strong>6100CR</strong> <strong>Refrigerated</strong> <strong>Sampler</strong><br />

Section 3 Installation of the <strong>Sampler</strong><br />

11. Reinsert the valve stem into the valve body by pressing<br />

and twisting the valve stem into the valve body at the<br />

same time. Turn the valve stems so their final position is<br />

perpendicular (90°) to the bottle.<br />

12. Reassemble the bottle rack with the cleaned bottles. Make<br />

sure all valves are closed. It is very important to keep the<br />

valve stems closed to prevent contamination.<br />

CAUTION<br />

The bottles used in the <strong>6100CR</strong> are borosilicate (hard) glass,<br />

also known by the trade name of Pyrex ® . Their capacity is 40<br />

ml. Teledyne <strong>Isco</strong> does not recommend the use of bottles other<br />

than those supplied by <strong>Isco</strong> to ensure conformity with the original<br />

equipment dimensions. If you use bottles from a source<br />

other than <strong>Isco</strong>, the bottles must have identical length, diameter,<br />

and cap thread as those supplied with the sampler. For<br />

safety reasons they should also be made of hard glass. Failure<br />

to use identical bottles may result in crushed bottles or a damaged<br />

needle assembly, and will result in an inoperable sampler.<br />

3.5 The Booster Pump Teledyne <strong>Isco</strong> offers a booster pump accessory for <strong>6100CR</strong> samplers<br />

when the pressure from the incoming sample lines is found<br />

to be inadequate to properly supply the sampler. Minimum<br />

sample line incoming pressure must be at least 20 psig. The<br />

booster pump raises the pressure of the incoming line to 35 psig<br />

if it is below the 20 psig minimum. This is enough pressure to<br />

allow accurate calibration of the sampler at 2 ml/sec.<br />

The booster pump consists of a housing, controls, and a<br />

motor-driven gear pump. To maintain the integrity of the<br />

incoming sample fluid, the pump gears are made of Teflon. The<br />

housing containing the booster pump is mounted on the side of<br />

the <strong>6100CR</strong> (see Figure 3-8). The interior of the booster pump<br />

housing is shown in the Figure 3-9.<br />

Note<br />

The pump must never be run without liquid inside the lines.<br />

Operation of the pump without liquid will destroy the Teflon<br />

gears inside the pump head.<br />

3-11