Section 4 Maintenance and Special Features<br />

2. The needle assembly consists of the needle, a heavy coil<br />

spring, three long slide rods, and four plastic rings. The<br />

plastic rings are knurled (lines on their surface). These<br />

rings are the needle mounts.<br />

3. To remove the needle assembly, loosen the guide nut, the<br />

top gray plastic nut in the middle of the assembly.<br />

4. Remove the drain hose from the bottom of the assembly.<br />

5. Push up on the ball on the lower end of the needle assembly<br />

to compress the ball and remove the top remaining nut.<br />

Note<br />

There is spring pressure on the assembly forcing it down. Hold<br />

the assembly up while removing the nut, then release slowly to<br />

prevent damage to the needle assembly.<br />

When pressure is released, disconnect the sample tube<br />

from the top of the needle and remove the assembly from<br />

the tower.<br />

6. Remove the needle from the assembly by removing the<br />

three screws holding the guide rods to the needle on the<br />

bottom. DO NOT loosen the retaining nuts on the top end<br />

of the slide rods. Pliers may be required to hold the slide<br />

rods while removing the screws. Remove the slide rods and<br />

unscrew the needle from the top needle mount, then<br />

remove the guide and spring.<br />

7. Reassemble in the reverse order. Make sure the needle<br />

assembly is seated against the top mount before tightening<br />

the nut. Align the drain tube fitting with the hose slot and<br />

make sure the slide rods are not twisted. Tighten the gray<br />

guide nut in the middle of the needle assembly until it contacts<br />

the mount. The assembly should still be able to move<br />

with light pressure applied to it.<br />

Note<br />

When reassembling the needle and tower, do not use tools to<br />

tighten the plastic mounting rings. You only need to finger-tighten<br />

them. You can use pliers to tighten the needle into<br />

the top mounting ring, but do not apply excessive force.<br />

4-2