Discrete input wiring connections<br />

W A R N I N G<br />

Figure 9 Discrete input wiring connections<br />

Potential Electrocution Hazard. Always disconnect power to the<br />

instrument when making electrical connections.<br />

W A R N I N G<br />

Potential electrocution hazard. In order to maintain the NEMA/IP<br />

environmental ratings of the enclosure, use only conduit fittings and<br />

cable glands rated for at least NEMA 4X/IP66 to route cables in to the<br />

instrument.<br />

Three discrete inputs are provided for switch closure inputs or logic level<br />

voltage inputs. Make wiring connections and configure jumper settings to<br />

the controller as shown in Figure 9, Table 4 and Figure 10.<br />

Note: Figure 9 shows the back of the controller cover and not the inside of the<br />

main controller compartment.<br />

Table 4 Input connections<br />

Discrete inputs Connector position -<br />

Switch input<br />

Connector position -<br />

Voltage input<br />

Input 1+ 3 2<br />

Input 1- 2 3<br />

Input 2+ 6 5<br />

English 19