Logarithmic output mode<br />
Figure 15 shows in graph form the operation of the logarithmic output<br />
mode.<br />
Figure 16 Bilinear output<br />
Figure 15 Logarithmic output<br />
1 Output current axis 5 Low value<br />
2 Source value axis 6 Minimum output current (0-4 mA)<br />
1 Output current axis 5 Minimum output current (0-4 mA)<br />
2 Source value axis 6 50% output current<br />
3 High value 7 Maximum output current (20 mA)<br />
4 50% value<br />
Bilinear output mode<br />
Figure 16 shows in graph form the operation of the bilinear output mode.<br />
3 High value 7 Knee point current<br />
4 Knee point value 8 Maximum output current (20 mA)<br />
Configure relays<br />
The Normally Open (NO) and Common (COM) relay contacts will be<br />
connected when an alarm or other condition is active. The Normally<br />
Closed (NC) and Common relay contacts will be connected when an<br />
alarm or other condition is inactive (unless the Fail Safe is set to Yes), or<br />
when power is removed from the controller. To select a menu option,<br />
highlight the option and push ENTER.<br />
1. From the <strong>sc200</strong> Setup menu, select Relay Setup.<br />
2. Select a relay from the list.<br />
28 English