• Alarm Function (refer to Figure 18)<br />
Figure 18 Alarm function<br />
Option<br />
Low alarm<br />
High alarm<br />
Low<br />
deadband<br />
High<br />
deadband<br />
Off delay<br />
On delay<br />
Description<br />
Sets the value where the relay will turn on in response to<br />
decreasing measured value. For example, if the low alarm is<br />
set for 1.0 and the measured value drops to 0.9, the relay<br />
activates.<br />
Sets the value where the relay will turn on in response to<br />
increasing measured value. For example, if the high alarm is<br />
set for 1.0 and the measured value increases to 1.1, the<br />
relay activates.<br />
Sets the range where the relay remains on after the<br />
measured value increases above the low alarm value. For<br />
example, if the low alarm is set for 1.0 and the low<br />
deadband is set for 0.5, the relay remains on between<br />
1.0 and 1.5. Default is 5% of the range.<br />
Sets the range where the relay remains on after the<br />
measured value decreases below the high alarm value. For<br />
example, if the high alarm is set for 4.0 and the high<br />
deadband is set for 0.5, the relay remains on between<br />
3.5 and 4.0. Default is 5% of the range.<br />
Sets a time (0-300 seconds) to delay the relay from normally<br />
turning off (Default: 0 seconds).<br />
Sets a delay time for the relay to turn on (Default:<br />
0 seconds).<br />
1 High alarm 5 ON delay<br />
2 High deadband 6 OFF delay<br />
3 Low deadband 7 Time (x-axis)<br />
4 Low alarm 8 Source (y-axis)<br />
30 English