Rosemount 3051S Electronic Remote Sensors - Emerson Process ...
Rosemount 3051S Electronic Remote Sensors - Emerson Process ...
Rosemount 3051S Electronic Remote Sensors - Emerson Process ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Reference Manual<br />
00809-0100-4804, Rev AA<br />
January 2011<br />
ROSEMOUNT<br />
MANIFOLDS<br />
Figure 2-14. <strong>Rosemount</strong> 305<br />
Integral Manifolds<br />
Figure 2-15. <strong>Rosemount</strong> 304<br />
Conventional Manifolds<br />
<strong>Rosemount</strong> <strong>3051S</strong> ERS System<br />
Transmitter Case<br />
Always ground the transmitter case in accordance with national and local<br />
electrical codes. The most effective transmitter case grounding method is a<br />
direct connection to earth ground with minimal impedance (< 1 ohm).<br />
Methods for grounding the transmitter case include:<br />
• Internal Ground Connection: The Internal Ground Connection screw is<br />
inside the terminal side of the electronics housing. The screw is<br />
identified by a ground symbol ( ), and is standard on all <strong>3051S</strong><br />
ERS transmitters.<br />
• External Ground Connection: The External Ground Connection is on<br />
the outside of the SuperModule housing. The connection is identified<br />
by a ground symbol ( ).<br />
The <strong>Rosemount</strong> 305 Integral Manifold assembles directly to an ERS sensor,<br />
eliminating the need for the flange. The 305 is available in two designs:<br />
Coplanar (bottom process connections) and Traditional (side process<br />
connections).<br />
305 COPLANAR<br />
The <strong>Rosemount</strong> 304 Conventional Manifold assembles directly to an<br />
instrument flange for easy servicing and retrofitting. The 304 manifold is<br />
available in two basic styles: Traditional (flange x flange and flange x pipe)<br />
and Wafer.<br />
304 TRADITIONAL 304 WAFER<br />
305 TRADITIONAL<br />
The <strong>Rosemount</strong> 306 Manifold assembles directly to an In-Line style<br />
transmitter and is available with male or female 1 /2-in. NPT process<br />
connections.<br />
2-21