Reference Manual<br />

00809-0100-4804, Rev AA<br />

January 2011<br />

<strong>Process</strong> Temperature<br />

Limits<br />

<strong>Rosemount</strong> <strong>3051S</strong> ERS System<br />

Storage<br />

• -50 to 185 °F (-46 to 85 °C)<br />

• With LCD display: -40 to 185 °F (-40 to 85 °C)<br />

At 70 °F (21 °C) ambient temperature and atmospheric pressure or above<br />

Table A-3. Gage / Absolute Sensor<br />

Coplanar Gage / Absolute Sensor<br />

In-Line Gage Sensor / Absolute Sensor<br />

Configuration<br />

(<strong>3051S</strong>AM_ _G, <strong>3051S</strong>AM_ _A)<br />

(<strong>3051S</strong>AM_ _T, <strong>3051S</strong>AM_ _E)<br />

Silicone Fill Fluid (1)<br />

- -40 to 250 °F (-40 to 121 °C) (2)<br />

with Coplanar Flange -40 to 250 °F (-40 to 121 °C) (2) -<br />

with Traditional Flange -40 to 300 °F (-40 to 149 °C) (2) -<br />

with Level Flange -40 to 300 °F (-40 to 149 °C) (2)<br />

-<br />

with 305 Integral Manifold -40 to 300 °F (-40 to 149 °C) (2) -<br />

Inert Fill Fluid (1)(3) -40 to 185 °F (-40 to 85 °C) (4) -22 to 250 °F (-30 to 121 °C) (2)<br />

(1) <strong>Process</strong> temperatures above 185 °F (85 °C) require de-rating the ambient limits by a 1.5:1 ratio. For example, for process temperature of 195 °F (91 °C),<br />

new ambient temperature limit is equal to 170 °F (77 °C). This can be determined as follows: (195 °F - 185 °F) x 1.5 = 15 °F, 185 °F - 15 °F = 170 °F.<br />

(2) 220 °F (104 °C) limit in vacuum service; 130 °F (54 °C) for pressures below 0.5 psia.<br />

(3) Not available with <strong>3051S</strong>AM_ _A<br />

(4) 160 °F (71 °C) limit in vacuum service.<br />

Table A-4. Liquid Level Sensor (<strong>3051S</strong>AL)<br />

<strong>Process</strong> Temperature Limits at 70 °F (21 °C) Ambient and Atmospheric Pressure or Above<br />

Fill Fluid<br />

No Extension 2-in (50 mm) Extension 4-in (100 mm) Extension Thermal Optimizer (1)<br />

200 Silicone -49 to 401 °F<br />

-49 to 401 °F<br />

-49 to 401 °F<br />

-49 to 401 °F<br />

(-45 to 205 °C)<br />

(-45 to 205 °C)<br />

(-45 to 205 °C)<br />

(-45 to 205 °C)<br />

704 Silicone 32 to 401 °F<br />

32 to 464 °F<br />

32 to 500 °F<br />

32 to 599 °F<br />

(0 to 205 °C)<br />

(0 to 240 °C)<br />

(0 to 260 °C)<br />

(0 to 315 °C)<br />

Inert (Halocarbon) -49 to 320 °F<br />

-49 to 320 °F<br />

-49 to 320 °F<br />

-49 to 320 °F<br />

(-45 to 160 °C)<br />

(-45 to 160 °C)<br />

(-45 to 160 °C)<br />

(-45 to 160 °C)<br />

Syltherm ® XLT Silicone -102 to 293 °F<br />

-102 to 293 °F<br />

-102 to 293 °F<br />

-102 to 293 °F<br />

(-75 to 145 °C)<br />

(-75 to 145 °C)<br />

(-75 to 145 °C)<br />

(-75 to 145 °C)<br />

Glycerin and Water 5 to 203 °F<br />

5 to 203 °F<br />

5 to 203 °F<br />

5 to 203 °F<br />

(-15 to 95 °C)<br />

(-15 to 95 °C)<br />

(-15 to 95 °C)<br />

(-15 to 95 °C)<br />

Propylene Glycol and<br />

5 to 203 °F<br />

5 to 203 °F<br />

5 to 203 °F<br />

5 to 203 °F<br />

Water<br />

(-15 to 95 °C)<br />

(-15 to 95 °C)<br />

(-15 to 95 °C)<br />

(-15 to 95 °C)<br />

Neobee M-20 5 to 401 °F<br />

5 to 437 °F<br />

5 to 437 °F<br />

5 to 437 °F<br />

(-15 to 205 °C)<br />

(-15 to 225 °C)<br />

(-15 to 225 °C)<br />

(-15 to 225 °C)<br />

(1) Consult Figure A-1 below for the specific operating range of using the Thermal Optimizer with 704 Silicone Fill Fluid.<br />