DP Calibration The DP calibration function can be used to adjust the calculated DP<br />
Fast Keys 3, 4, 2, 1<br />
measurement of the ERS system. For example, a DP zero trim can be<br />
performed if the calculated DP of the ERS system has a small offset when the<br />
expected output should be “0 DP.”<br />
NOTE<br />
Because the DP calculation is dependent on the P HI and P LO pressure<br />
measurements, all DP calibration functions should be performed after<br />
completing the calibration functions on the individual P HI and P LO sensors.<br />
While a lower or upper DP trim is not required, a zero DP trim should be<br />
performed after installing and calibrating the individual pressure sensors and<br />
before subjecting the ERS system to the actual process conditions in order to<br />
establish a zero-based DP measurement.<br />
DP Zero Trim<br />
The DP Zero Trim function establishes a true zero-based DP calculation by<br />
taking the current measurement output and forcing that value as the new<br />
zero-reference. A DP Zero Trim should only be performed when the expected<br />
output of the ERS system is “0 DP.” For non-zero based trims, a DP Lower<br />
Trim should be performed instead.<br />
The DP zero trim function requires both pressure sensors to be wired and<br />
connected.<br />
To perform a DP zero trim, follow the procedure below:<br />
1. Ensure that the individual PHI and PLO pressure sensors have been<br />
calibrated as detailed on page 4-3 and are wired together as shown in<br />
Figure 2-9, Figure 2-10, or Figure 2-11.<br />
2. Launch the “DP Zero Trim” function using AMS or a Field Communicator.<br />
3. Apply “0 DP” to the ERS System and wait for the DP measurement to<br />
stabilize.<br />
4. Using AMS or a Field Communicator, perform the zero trim function on<br />
the <strong>3051S</strong> ERS System.<br />
Upper and Lower DP Trims<br />
The DP calculation can be trimmed using a two-point calibration where lower<br />
and upper end-point pressures are applied, and all readings are linearized<br />
between the two points.<br />
Unlike the DP Zero Trim function, the Upper and Lower DP Trims can be<br />
performed when the ERS system is pressurized under actual process<br />
conditions.<br />
Always perform a lower DP trim first to establish the correct offset. The upper<br />
DP trim provides a slope correction.<br />