Rosemount 3051S Electronic Remote Sensors - Emerson Process ...
Rosemount 3051S Electronic Remote Sensors - Emerson Process ...
Rosemount 3051S Electronic Remote Sensors - Emerson Process ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Reference Manual<br />
00809-0100-4804, Rev AA<br />
January 2011 <strong>Rosemount</strong> <strong>3051S</strong> ERS System<br />
Appendix A Specifications and<br />
Reference Data<br />
PERFORMANCE<br />
SPECIFICATIONS<br />
Conformance to<br />
Specification (±3�)<br />
Reference Accuracy<br />
Table A-1. DP Reference Accuracy of <strong>3051S</strong> ERS System (1)<br />
Two Coplanar Gage <strong>Sensors</strong> (<strong>3051S</strong>AM_ _G)<br />
www.rosemount.com<br />
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page A-1<br />
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page A-4<br />
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-9<br />
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-11<br />
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-17<br />
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-37<br />
For zero-based spans, reference conditions, silicone oil fill, glass-filled PTFE<br />
o-rings, SST materials, Coplanar flange, or 1 /2 in.-14 NPT process<br />
connections, digital trim values are set to equal range points.<br />
Technology leadership, advanced manufacturing techniques, and statistical<br />
process control ensure specification conformance to at least ±3�.<br />
Ultra Classic<br />
Range 2-4 ±0.035% of DP span ±0.078% of DP span<br />
Range 5 ±0.071% of DP span ±0.092% of DP span<br />
Two Coplanar Absolute <strong>Sensors</strong> (<strong>3051S</strong>AM_ _A)<br />
Ultra Classic<br />
Ranges 1-4 ±0.035% of DP span ±0.078% of DP span<br />
Two In-Line Gage <strong>Sensors</strong> (<strong>3051S</strong>AM_ _T)<br />
Two In-Line Absolute <strong>Sensors</strong> (<strong>3051S</strong>AM _ _E)<br />
Ultra Classic<br />
Ranges 1-4 ±0.035% of DP span ±0.078% of DP span<br />
Two Liquid Level <strong>Sensors</strong> (<strong>3051S</strong>AL)<br />
Ultra Classic<br />
Ranges 1-5 ±0.092% of DP span ±0.092% of DP span<br />
(1) DP Accuracy specifications for the <strong>3051S</strong> ERS System assume that the configuration contains two sensors with identical ranges that are calibrated<br />
0 – URL, and that the DP Span = 10% of each sensor’s URL.