12.2 X-ray device settingsCheck the following appliance-specific standard values and adapt as required.Reference conditions:• Adult patientParameterDC-emitter, 7 mA,tube length 20 cmDC-emitter, 7 mA,tube length 30 cmENrecommendedexposure time (s)recommendedexposure time (s)upper jaw 60kV 70kV 60kV 70kVincisors 0.1 0.08 0.2 0.16pre-molars 0,125 0.1 0.25 0.2molars 0.16 0,125 0.32 0.25lower jaw 60kV 70kV 60kV 70kVincisors 0.1 0.08 0.2 0.16pre-molars 0,125 0.1 0.25 0.2molars 0,125 0.1 0.25 0.2Bitewing 0.16 0,125 0.32 0.25Where it is possible to set the x-ray tube to 60 kV, this setting should be selected.The usual exposure values for F-Film (e.g. Kodak Insight) can be used.12.3. Commissioning and Constancy Checks*Before the appliance can be used the first time a handover test (commissioning test) of theX-ray system must be carried out according to national and local regulations. The test body forIntra / Extra (order number 2121-060-52) will be required.The Constancy Check of the X-ray system must be carried out by surgery staff at frequentintervals. The Constancy Check is based on those results obtained during the handover test.* observe all rules and regulations!18