19.4. Storage box with image platestorage trayStorage boxWhere the surface shows signs of contaminationor visible dirt, clean and disinfect the unit.• Clean the surfaces using a damp cloth.For disinfection, use a disinfectant spray(e.g. Dürr FD 366 sensitiv) on a soft, fluff-freecloth.Image plate storage trayWhere the surface shows signs of contaminationor visible dirt, clean and disinfect the unit.ENThe surface can be cleaned using a damp cloth,and then disinfected using a disinfection wipe(e.g. Dürr Dental FD 350).Additional the image plate storage tray could bedisinfected/sterilised with• Thermal disinfector• Autoclave.Do not exceed a temperatureof 134 °C. 31