R&S <strong>ZVL</strong>-<strong>K1</strong> Selecting <strong>the</strong> Frequency and Span – CENTER Key<br />

Softkeys of <strong>the</strong> frequency menu<br />

The following table shows all softkeys available in <strong>the</strong> frequency menu. It is possible that your<br />

instrument configuration does not provide all softkeys. If a softkey is only available with a special option,<br />

model or (measurement) mode, this in<strong>for</strong>mation is delivered in <strong>the</strong> corresponding softkey description.<br />

Menu / Command Command<br />

Center<br />

Start<br />

Stop<br />

CF Stepsize � 0.1*Span/0.1*RBW<br />

Frequency Offset<br />

0.5*Span/0.5*RBW<br />

x*Span/x*RBW<br />

=Center<br />

=Marker<br />

<strong>Manual</strong><br />

Signal Track � Track On/Off<br />

Center<br />

Start<br />

Track BW<br />

Track Threshold<br />

Select Trace<br />

Opens an edit dialog box to enter <strong>the</strong> center frequency. The allowed range of values <strong>for</strong> <strong>the</strong><br />

center frequency depends on <strong>the</strong> frequency span.<br />

span > 0: spanmin / 2 ≤ fcenter ≤ fmax – spanmin / 2<br />

span = 0: 0 Hz ≤ fcenter ≤ fmax<br />

fmax and spanmin are specified in <strong>the</strong> data sheet. To help analyze signals located at <strong>the</strong> end of <strong>the</strong><br />

frequency range, <strong>for</strong> R&S <strong>ZVL</strong>-<strong>K1</strong> models with an upper frequency limit of 6 GHz or less, <strong>the</strong> fmax<br />

value is extended by 0.05 GHz <strong>for</strong> direct entry via <strong>the</strong> key pad. The preset and full span values<br />

remain unchanged.<br />

Remote: FREQ:CENT 100MHz<br />

Opens an edit dialog box to define <strong>the</strong> start frequency. The following range of values is allowed:<br />

fmin ≤ fstart ≤ fmax – spanmin<br />

fmin, fmax and spanmin are specified in <strong>the</strong> data sheet. To help analyze signals located at <strong>the</strong> end of<br />

<strong>the</strong> frequency range, <strong>for</strong> R&S <strong>ZVL</strong>-<strong>K1</strong> models with an upper frequency limit of 6 GHz or less, <strong>the</strong><br />

fmax value is extended by 0.05 GHz <strong>for</strong> direct entry via <strong>the</strong> key pad. The preset and full span<br />

values remain unchanged.<br />

� Remote: FREQ:STAR 20MHz<br />

1303.6573.42 3.7 E-5