28.12.2012 Views

Software Manual for the R&S ZVL-K1 - Rohde & Schwarz

Software Manual for the R&S ZVL-K1 - Rohde & Schwarz

Software Manual for the R&S ZVL-K1 - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>ZVL</strong>‑<strong>K1</strong> INITiate Subsystem<br />

INITiate Subsystem<br />

The INITiate subsystem is used to control <strong>the</strong> init–measurement function.<br />

Commands of <strong>the</strong> INITiate Subsystem<br />

– INITiate[:IMMediate]<br />

– INITiate:CONMeas<br />

– INITiate:CONTinuous<br />

– INITiate:DISPlay<br />

– INITiate:ESPectrum<br />

– INITiate:SPURious<br />

INITiate[:IMMediate]<br />

The command initiates a new measurement sequence.<br />

With sweep count > 0 or average count > 0, this means a restart of <strong>the</strong> indicated number of<br />

measurements. With trace functions MAXHold, MINHold and AVERage, <strong>the</strong> previous results are<br />

reset on restarting <strong>the</strong> measurement.<br />

In single sweep mode, synchronization to <strong>the</strong> end of <strong>the</strong> indicated number of measurements can<br />

be achieved with <strong>the</strong> command *OPC, *OPC? or *WAI. In continuous–sweep mode,<br />

synchronization to <strong>the</strong> sweep end is not possible since <strong>the</strong> overall measurement never ends.<br />

The numeric suffixes are irrelevant <strong>for</strong> this command.<br />

This command is an event and <strong>the</strong>re<strong>for</strong>e has no *RST value and no query.<br />

Example<br />

INIT:CONT OFF<br />

Switches to single sweep mode.<br />

DISP:WIND:TRAC:MODE AVER<br />

Activates trace averaging.<br />

SWE:COUN 20<br />

Setting <strong>the</strong> sweep counter to 20 sweeps.<br />

INIT;*WAI<br />

Starts <strong>the</strong> measurement and waits <strong>for</strong> <strong>the</strong> end of <strong>the</strong> 20 sweeps.<br />

Characteristics<br />

*RST value: –<br />

SCPI: con<strong>for</strong>m<br />

Mode<br />

all<br />

INITiate:CONMeas<br />

This command continues a stopped measurement at <strong>the</strong> current position in single sweep mode.<br />

The function is useful especially <strong>for</strong> trace functions MAXHold, MINHold and AVERage, if <strong>the</strong><br />

previous results are not to be cleared with sweep count > 0 or average count > 0 on restarting<br />

<strong>the</strong> measurement (INIT:IMMediate resets <strong>the</strong> previous results on restarting <strong>the</strong> measurement).<br />

The single sweep mode is automatically switched on. Synchronization to <strong>the</strong> end of <strong>the</strong><br />

indicated number of measurements can <strong>the</strong>n be per<strong>for</strong>med with <strong>the</strong> commands *OPC, *OPC? or<br />

1303.6573.42 4.143 E-5