R&S <strong>ZVL</strong>‑<strong>K1</strong> CALCulate Subsystem<br />

Characteristics<br />

RST value: OFF<br />

SCPI: device–specific<br />

Mode<br />

all<br />

CALCulate:DELTamarker:AOFF<br />

This command switches off all active delta markers.<br />

For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) <strong>the</strong> numeric<br />

suffix selects <strong>the</strong> measurement screen. In <strong>the</strong> o<strong>the</strong>r modes <strong>the</strong> numeric suffix is<br />

irrelevant.<br />

Example<br />

CALC:DELT:AOFF<br />

Switches off all delta markers.<br />

Characteristics<br />

RST value: –<br />

SCPI: device–specific<br />

Mode<br />

A, ADEMOD, CATV, CDMA, EVDO, WCDMA<br />

CALCulate:DELTamarker:FUNCtion:FIXed[:STATe]<br />

This command switches <strong>the</strong> relative measurement to a fixed reference value on or off. Marker 1<br />

will be activated previously and a peak search will be per<strong>for</strong>med, if necessary. If marker 1 is<br />

activated, its position becomes <strong>the</strong> reference point <strong>for</strong> <strong>the</strong> measurement. The reference point<br />

can <strong>the</strong>n be modified with <strong>the</strong><br />

CALCulate:DELTamarker:FUNCtion:FIXed:RPOint:X commands and<br />

CALCulate:DELTamarker:FUNCtion:FIXed:RPOint:Y independently of<br />

<strong>the</strong> position of marker 1 and of a trace. It applies to all delta markers as long as <strong>the</strong> function is<br />

active.<br />

The numeric suffixes are irrelevant <strong>for</strong> this command.<br />

Parameter<br />

ON | OFF<br />

Example<br />

CALC:DELT:FUNC:FIX ON<br />

Switches on <strong>the</strong> measurement with fixed reference value <strong>for</strong> all delta markers.<br />

CALC:DELT:FUNC:FIX:RPO:X 128 MHZ<br />

Sets <strong>the</strong> frequency reference to 128 MHz.<br />

CALC:DELT:FUNC:FIX:RPO:Y 30 DBM<br />

Sets <strong>the</strong> reference level to +30 dBm.<br />

Characteristics<br />

RST value: OFF<br />

SCPI: device–specific<br />

Mode<br />

A<br />

1303.6573.42 4.13 E-5

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!