<strong>Directions</strong> <strong>for</strong> <strong>Use</strong> Introduction 5<br />

Blood Pressure Warnings<br />

WARNING Do not autoclave.<br />

WARNING This device complies with current required standards <strong>for</strong><br />

electromagnetic interference and should not present problems to other<br />

equipment or be affected by other devices. As a precaution, avoid using this<br />

device in close proximity to other equipment.<br />

WARNING Welch Allyn is not responsible <strong>for</strong> the integrity of any mounting<br />

installation. Welch Allyn recommends that the customer contact their Biomedical<br />

Engineering Department or maintenance service to ensure professional<br />

installation <strong>for</strong> safety and reliability of any mounting accessory.<br />

WARNING The <strong>Spot</strong> <strong>LXi</strong> consists of high-quality precision parts. Protect it from<br />

severe impact and shock. A qualified service technician must check any <strong>Spot</strong> <strong>LXi</strong><br />

that is dropped or damaged <strong>for</strong> proper operation prior to further use. Do not use<br />

the <strong>Spot</strong> <strong>LXi</strong> if you notice any signs of damage. Contact the Welch Allyn<br />

Customer Service Department <strong>for</strong> assistance.<br />

WARNING Do not use an SpO 2 finger clip sensor and a blood pressure cuff<br />

simultaneously on the same limb. Doing so may result in inaccurate pulse rate<br />

and perfusion readings.<br />

WARNING The USB port is <strong>for</strong> service only. Do not use this connection <strong>for</strong> any<br />

external equipment connections.<br />

These warnings pertain to the <strong>Spot</strong> <strong>LXi</strong> blood pressure feature.<br />

WARNING <strong>Spot</strong> <strong>LXi</strong> is not intended to measure BLOOD PRESSURE on<br />

neonatal patients. The AAMI SP10:2002 standard defines neonates as children 28<br />

days or less of age if born at term (37 weeks gestation or more); otherwise up to<br />

44 gestational weeks.<br />

WARNING To ensure pediatric blood pressure accuracy and safety, the Child<br />

Reusable Two-Piece Blood Pressure Cuff (4500-01), Infant Durable One-Piece<br />

Cuff (5082-82-4MQ), and the Infant Disposable One-Piece Cuff (5082-92-4MQ)<br />

are the smallest cuffs approved <strong>for</strong> use with young children and infants. The<br />

child’s arm must fit within the range markings on the cuff.<br />

WARNING Avoid compression of the blood pressure hose or cuff tubing of <strong>Spot</strong><br />

<strong>LXi</strong>. This may cause system errors to occur in the device.<br />

WARNING Patients who are experiencing moderate to severe arrhythmias may<br />

give inaccurate blood pressure measurements.<br />

WARNING <strong>Spot</strong> <strong>LXi</strong> does not operate effectively on patients who are<br />

experiencing convulsions or tremors.<br />

WARNING <strong>Use</strong> only Welch Allyn blood pressure cuffs and/or hoses. Using other<br />

manufacturers’ blood pressure cuffs and/or hoses may produce inaccurate blood<br />

pressure readings.<br />

WARNING When several blood pressure measurements are taken on the same<br />

patient, regularly check the cuff site and extremity <strong>for</strong> possible ischemia, purpura,<br />

and/or neuropathy.