Blood Pressure Warnings<br />
WARNING Do not autoclave.<br />
WARNING This device complies with current required standards <strong>for</strong><br />
electromagnetic interference and should not present problems to other<br />
equipment or be affected by other devices. As a precaution, avoid using this<br />
device in close proximity to other equipment.<br />
WARNING Welch Allyn is not responsible <strong>for</strong> the integrity of any mounting<br />
installation. Welch Allyn recommends that the customer contact their Biomedical<br />
Engineering Department or maintenance service to ensure professional<br />
installation <strong>for</strong> safety and reliability of any mounting accessory.<br />
WARNING The <strong>Spot</strong> <strong>LXi</strong> consists of high-quality precision parts. Protect it from<br />
severe impact and shock. A qualified service technician must check any <strong>Spot</strong> <strong>LXi</strong><br />
that is dropped or damaged <strong>for</strong> proper operation prior to further use. Do not use<br />
the <strong>Spot</strong> <strong>LXi</strong> if you notice any signs of damage. Contact the Welch Allyn<br />
Customer Service Department <strong>for</strong> assistance.<br />
WARNING Do not use an SpO 2 finger clip sensor and a blood pressure cuff<br />
simultaneously on the same limb. Doing so may result in inaccurate pulse rate<br />
and perfusion readings.<br />
WARNING The USB port is <strong>for</strong> service only. Do not use this connection <strong>for</strong> any<br />
external equipment connections.<br />
These warnings pertain to the <strong>Spot</strong> <strong>LXi</strong> blood pressure feature.<br />
WARNING <strong>Spot</strong> <strong>LXi</strong> is not intended to measure BLOOD PRESSURE on<br />
neonatal patients. The AAMI SP10:2002 standard defines neonates as children 28<br />
days or less of age if born at term (37 weeks gestation or more); otherwise up to<br />
44 gestational weeks.<br />
WARNING To ensure pediatric blood pressure accuracy and safety, the Child<br />
Reusable Two-Piece Blood Pressure Cuff (4500-01), Infant Durable One-Piece<br />
Cuff (5082-82-4MQ), and the Infant Disposable One-Piece Cuff (5082-92-4MQ)<br />
are the smallest cuffs approved <strong>for</strong> use with young children and infants. The<br />
child’s arm must fit within the range markings on the cuff.<br />
WARNING Avoid compression of the blood pressure hose or cuff tubing of <strong>Spot</strong><br />
<strong>LXi</strong>. This may cause system errors to occur in the device.<br />
WARNING Patients who are experiencing moderate to severe arrhythmias may<br />
give inaccurate blood pressure measurements.<br />
WARNING <strong>Spot</strong> <strong>LXi</strong> does not operate effectively on patients who are<br />
experiencing convulsions or tremors.<br />
WARNING <strong>Use</strong> only Welch Allyn blood pressure cuffs and/or hoses. Using other<br />
manufacturers’ blood pressure cuffs and/or hoses may produce inaccurate blood<br />
pressure readings.<br />
WARNING When several blood pressure measurements are taken on the same<br />
patient, regularly check the cuff site and extremity <strong>for</strong> possible ischemia, purpura,<br />
and/or neuropathy.