8<br />

Error Codes<br />

Troubleshooting<br />

The following tables of error codes provide a quick reference of the descriptions and<br />

probable causes of error codes. For service-level troubleshooting, refer to the Service<br />

Manual (part number 704432).<br />

Table 12. General Error Codes<br />

Code Description Corrective Action<br />

C12 Device outside operating temperature<br />

range.<br />

Change ambient temperature.<br />

C13 Low battery level. Charge battery.<br />

E30 Internal malfunction. Contact Technical Service.<br />

E31 Internal malfunction. Contact Technical Service.<br />

E32 Internal malfunction. Contact Technical Service.<br />

E33 Internal malfunction. Contact Technical Service.<br />

E38 Date and time not set. Set date and time (see “Date/Time Menu Options” on<br />

page 32)<br />

E42 Internal malfunction. Contact Technical Service.<br />

E44 Internal malfunction. Contact Technical Service.<br />

E45 Internal malfunction. Contact Technical Service.<br />

Table 13. Blood Pressure Error Codes<br />

Code Description Corrective Action<br />

C01 Blood pressure reading cancelled by user. Retake blood pressure reading.<br />

C02 Unable to release cuff pressure. Check tubing <strong>for</strong> kinks and connection integrity.<br />

C03 Inflation too quick. Check tubing and connections.<br />

C04 Air leak. Check blood pressure cuff and tubing connections.<br />

C05 Unable to determine blood pressure. Check connections; restrict patient movement.<br />

C06 Unable to determine blood pressure. Check connections; restrict patient movement.<br />

C07 Internal NIBP error. Device will power down.<br />

E10 Cuff pressure limits exceeded. Device will power down.<br />

E11 Cuff pressure duration exceeded. Device will power down.<br />

E20 Internal NIBP error. Device will power down.<br />