44 Pulse Oximetry Operation Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> <strong>LXi</strong><br />
<strong>Spot</strong> <strong>LXi</strong> measures a patient’s SpO 2 <strong>for</strong> up to 10 minutes. After 10 minutes, a C9 error<br />
code is displayed. This error code means that the use has exceeded the 10-minute<br />
time limit.<br />
2. Check sensor sites periodically to assess circulation, sensor positioning, and skin<br />
sensitivity.<br />
3. Remove the sensor from the patient. The SpO 2 reading flashes <strong>for</strong> 8 seconds. If the<br />
sensor is not reattached to the patient in 8 seconds, the SpO 2 % and pulse rate<br />
flashing, signalling that the measurement period has ended. The device continues to<br />
display the last SpO 2 reading.