07.01.2014 Views

CHEM01200604009 Sreejith Kaniyankandy - Homi Bhabha ...

CHEM01200604009 Sreejith Kaniyankandy - Homi Bhabha ...

CHEM01200604009 Sreejith Kaniyankandy - Homi Bhabha ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

63<br />

The electrical signal is then channeled to CFD and subsequently to TAC, which stops<br />

charging. The TAC output is channeled to a MCP which gives a numerical value for the time<br />

delay through an analog to digital converter. The value associated is added to different<br />

channels corresponding to different delay times and a combined histogram is generated for<br />

each of the channels and the waveform obtained is the decay profile at that wavelength.<br />

Once the signal is detected it needs to be analyzed to delink lamp profile and different<br />

components in decay. If excitation pulse is a delta (δ)-pulse or sharp pulse and detection is<br />

very fast the decay profile is the actual decay of the system. However excitation pulse always<br />

has a finite width and detection system also has a finite response time. Therefore the decay<br />

profile measured is actually a convolution of decay profile of the samples and the light<br />

source. Hence mathematical deconvolution procedure is used to obtain the fits. To perform a<br />

deconvolution procedure we need to obtain the profile of the light source. This is done by<br />

using a scattering solution with no emission and measure Rayleigh scattering for that<br />

excitation wavelength. The profile hence obtained is called an instrument response function.<br />

The width contains information on profile of both detector and lamp profile. Decay profiles<br />

are generated with a fixed angle between excitation and emission polarization, this angle is<br />

called magic angle (54.7˚). This is done to avoid anisotropy effects in decay profiles<br />

generated. The maximum decay counts was taken as 5000 or 10000 for all the samples<br />

measured. This is done to ensure good statistics for fitting. TAC range used in the present<br />

thesis was set from 50-250ns depending on different systems studied in the present thesis.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!