48<br />
<strong>Laboratory</strong> <strong>Automation</strong> <strong>Hardware</strong><br />
4 Digital Inputs and 4 Digital Outputs and Feeding (LP-DE4DA4PA)<br />
This panel respectively combines<br />
four channels of the LP-DEP8<br />
and LP-DAP8A panels. See in the<br />
according section for detailed<br />
information.<br />
Technical Data<br />
Connection example for magnetic valves and fill level switches.<br />
Signal NAMUR compliant input level, active change-over contact (normally open contact<br />
+24 V, normally close contact +24 V), capable of bearing 1 A<br />
Connection type See LP-DEP8, LP-DAP8A<br />
Usable with HK-DEA16N, HS-NGS241A to HS-NGS245<br />
For connecting Switch contacts, fill level switches, push switches, valves, reflux divisors, motors,<br />
digital control inputs on devices etc.<br />
Delivery Scope Panel with holding bolts, cable for internal connection<br />
Order Code Description<br />
LP-DE4DA4PA Combination panel 4 digital inputs DEP8, 4 active digital outputs DAP8A<br />
LZ-AKTUB Caps for tuchel sockets<br />
LZ-AKTUS Caps for tuchel plugs<br />
For suitable connecting cables for valves, actuators etc. see "Cable for Instrumentation“.<br />
4 Digital Outputs, 4 Heating Elements and Feeding (LP-DA4HZ4PA)<br />
This panel combines four<br />
channels for low-voltage<br />
heatings with four channels of<br />
the panel.<br />
For details, see LP-DAP8(A).<br />
Technical Data<br />
Signal 44 V AC, max. 2,5 A per output, active change-over contact (normally open<br />
contact +24 V, normally close contact +24 V), capable of bearing 1 A<br />
Applicable with HK-DEA16N, HS-NGS44AC<br />
For the Connection of Switch Contacts, Liquid Level Switch , Pressure switches, Low Voltage Heatings<br />
Scope of Delivery Panel with fastening Screws, Cable for internal cabling<br />
Order Code Description<br />
LP-LP-DA4HZ4PA Combination Panel 4 digital Outputs 24 V 1 A, 4 Low Voltage Heatings<br />
LZ-AKTUB Covers for Tuchel-socket<br />
For suited connection cables for valves, actuators etc., see chapter "Cables for Instrumenting“.<br />
30 Years of Excellence in Lab Scale Technology · Over 1.000 Products around the <strong>Laboratory</strong> Reactor