100 Bear Cub 750PSV<br />

Table 9-8: Adult Ventilator Monitor Status<br />

Mode: Upon Request<br />

Length: 18 Bytes<br />

Format:<br />

[ SYN ] [ID] [Data #1] [Data #15] [Check]<br />

[ ID ] : 60h<br />

[ Data #1 ] Breath Data<br />

Bit 7 — Control Breath (1=on, 0=off)<br />

Bit 6 — Sigh Breath (1=on, 0=off)<br />

Bit 5 — Patient Effort (1=on, 0=off)<br />

Bit 4 — MMV Active (1=on, 0=off)<br />

Bit 3 — Reserved (0)<br />

Bit 2 — Reserved (0)<br />

Bit 1 — Reserved (0)<br />

Bit 0 — Reserved (0)<br />

[Data #n] Parameter Range Binary Scale Binary Range<br />

[Data #2] MSB Tidal Vol 0 - 9.99 L 0.01 L 0 - 999<br />

[Data #3] LSB " "<br />

[Data #4] MSB Minute Vol 0 - 99.9 L 0.1 L 0 - 999<br />

[Data #5] LSB " "<br />

[Data #6] MSB Spon Min Vol 0 - 99.9 L 0.1 L 0 - 999<br />

[Data #7] LSB " "<br />

[Data #8] MSB 1/I:E Ratio 0.00 - 4.0 0.01 0 - 400<br />

[Data #9] LSB " "<br />

[Data #10] Total Rate 0 - 155 bpm 1 bpm 0 - 155<br />

[Data #11] Spon Rate 0 - 155 bpm 1 bpm 0 - 155<br />

[Data #12] %MMV 0 - 100 %/hr 1 %/hr 0 - 100<br />

[Data #13] Mean Pres 0 - 140 cmH2O 1 cmH2O 0 - 140<br />

[Data #14] Peak Pres 0 - 140 cmH2O 1 cmH2O 0 - 140<br />

[Data #15] Plateau Pres 0 - 140 cmH2O 1 cmH2O 0 – 140<br />