19.11.2012 Views

Operator's Manual-AVEA - Static Content

Operator's Manual-AVEA - Static Content

Operator's Manual-AVEA - Static Content

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Service <strong>Manual</strong> 37<br />

Chapter 4 Calibration<br />

(for software releases 2 and 3 only)<br />

Introduction<br />

The information in this section is intended for use by a factory trained service<br />

technician. Before opening the console or removing assemblies to perform<br />

calibration, a reference should be made to the Instruction <strong>Manual</strong> and Sections 1, 2,<br />

3, and 7 of this <strong>Manual</strong> to assure that the ventilator is being properly utilized.<br />

WARNINGS<br />

• The calibration procedures contained in this section must be completed when<br />

any part/assembly has been removed and/or replaced. Failure to do so could<br />

result in ventilator malfunction and injury or death to the patient.<br />

• After calibration and/or replacement of any assembly(ies), always perform the<br />

Operational Verification Procedure (Section 3), then allow a "BURN-IN" period<br />

and repeat the Operational Verification Procedure.<br />

• Oxygen vigorously accelerates combustion. DO NOT use any gauges, valves or<br />

other equipment that has been exposed to oil or grease contamination to avoid<br />

violent ignition.<br />

• Hazardous voltages exist within the ventilator. Always observe appropriate<br />

safety precautions when working on the ventilator while the machine is<br />

connected to an electrical power source to prevent possible accidental injurious<br />

electrical shock.<br />

• When high pressure gas sources are connected to the ventilator, always use<br />

extreme caution when attempting to measure internal pressures. Abnormal<br />

conditions may exist where measured pressures exceed the burst pressure of<br />

the gauge, resulting in possible injury.<br />

NOTE<br />

The Operational Verification Procedure (OVP) is always performed prior to use of<br />

this section.<br />

CAUTION<br />

• Before using any test equipment (electronic or pneumatic) for calibration<br />

purposes (other than operational verification), the accuracy of the instruments<br />

must be certified by a testing laboratory. The laboratory master test instruments<br />

must be traceable to the US Bureau of Standards or equivalent. When variances<br />

exist between the indicated and actual values, the calibration curves (proved by<br />

the testing laboratory) must be used to establish the actual correct values. This<br />

certification procedure should be performed at least once every six months. More<br />

frequent certification may be required based on usage and environment.<br />

L2342 Revision B November 2004

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!