Service <strong>Manual</strong> 13<br />
Chapter 1: Specifications<br />
Modes<br />
A/C<br />
In the Assist/Control position, a mechanical breath is delivered with each inspiratory<br />
effort, provided the patient satisfies the assist sensitivity criteria. If the patient does<br />
not meet the assist sensitivity criteria, the ventilator automaticallydelivers breaths<br />
according to the Ventilator Rate control. Mechanical breaths may then be either<br />
patient initiated (assisted) or ventilator initiated (controlled).<br />
SIMV/IMV<br />
In the SIMV/IMV position, a combination of mechanical and spontaneous breath<br />
types is available. Mechanical breaths, either assisted or controlled, are delivered at<br />
the set ventilator rate while all other breaths are spontaneous. Disabling the assist<br />
sensitivity mechanism. or removing the flow sensor from the ventilator, will cause the<br />
equivalent of IMV mode of ventilation to exist.<br />
CPAP<br />
In the CPAP position, the Base Flow control establishes the continuous flow<br />
available for spontaneous breathing.<br />
STANDBY<br />
The STANDBY position discontinues all electrical controls and functions with the<br />
exception of the charging system for the internal battery.<br />
NOTE<br />
If the ventilator is connected to air and O2 sources, blended gas will circulate while<br />
ventilator is in the STANDBY position, making gas available from the Auxiliary Gas<br />
Outlet. The amount of gas flow circulating is determined by the Base Flow setting.<br />
Controls<br />
Rate<br />
Range: 1 to 150 BPM<br />
Resolution: 1 from 1 to 100BPM<br />
2 from 100 to 150 BPM<br />
Tolerance: Greater of ± 1 BPM or 0.02 seconds applied to the Breath<br />
Interval<br />
L2342 Revision B November 2004