Service <strong>Manual</strong> 33<br />

Verify that the audible alarm is canceled and the unit continues to function.<br />

Reconnect the AC power.<br />

21. Connect the circuit to the infant test lung. Set the Volume Limit to 25ml. Verify<br />

That the tidal Volume readout is 25 ml ± 20%.<br />

22. Disconnect the Flow Sensor cable from the ventilator. Verify that the flow<br />

sensor alarm activates, assist sensitivity window is blank, and that the<br />

monitored volume and volume limit displays show dashes. Reconnect the flow<br />

sensor cable, verify audible alarm cancels, assist sensitivity window<br />

illuminates, and volume digits return.<br />

23. Turn Assist Sensitivity to MIN. Squeeze and release the patient circuit. Verify<br />

that the unit will assist and that Patient Initiated LED is illuminating.<br />

24. Disconnect the patient circuit from the test lung. Verify that both audible and<br />

visual Low Inspiratory Pressure alarms are activated. Verify that the Alarm<br />

Loudness can be varied from minimum to maximum, but the alarm cannot be<br />

turned off.<br />

Figure 3-1: Bear Cub 750 US with Patient Circuit<br />

L2342 Revision B November 2004